DEP SOPs
 

Site index

Home

 

 

Created by:

 

11.0            CORRECTIVE ACTIONS

 

      Quality controls are used to monitor and assess the effectiveness and validity of a sampling or analysis activity.  If a specified quality control measure is determined to be out of a predetermined acceptance range, and the source or reason for the deviation is not identified and corrected, the sample data associated with the quality control measure may not be useful or valid information.

 

      Some quality control criteria (ex. calibration) have a direct effect on the test results.  Others (ex. blanks and duplicates) are indicators of improper protocols or contamination.

 

11.1            QUALITY CONTROL MEASURES AND ACCEPTANCE CRITERIA

 

      Table 11.1 identifies each of the quality control checks that are required by test methods and/or DER acceptance criteria.  The acceptance range criteria or the source of the acceptance range has been identified.

 

11.2            IDENTIFYING AND ASSESSING QC MEASURES

 

      Generally, quality control information is reviewed by several individuals.  The responsibility for the initial assessment of a quality control measure lies with the individual who (1) identifies the sample or procedure as a QC measure; and (2) has access to the test results:

 

11.2.1.  The individual responsible for operating the analytical instrument or equipment must be responsible for assessing the following applicable QC Measures:

      1.   Method, reagent and calibration blanks

      2.      Calibration integrity:  initial and continuing calibration, interference standards, and QC check standards

      3.   System performance checks

      4.   Tuning criteria

      5.      Surrogate and internal samples

      6.   Titrating solutions

 

11.2.2.  The following checks are normally assessed by a secondary reviewer (supervisor or QA Officer), but may be evaluated by the primary analyst:

      1.      Standard Reference Material

      2.   QC Check Samples

      3.   Spiked samples (matrix and blank)

      4.      Duplicates

 

11.2.3.  The following must be assessed by the organization or individual(s) responsible for sample collection, but may be reviewed by laboratory personnel if the sample has been identified as:

      1.      Precleaned and field cleaned equipment blanks

      2.   Trip blanks

      3.   Field collected duplicates

      4.   Split samples

 

11.3            DETERMINING THE SOURCE OF QC PROBLEMS

 

      Once a problem has been identified, the process (whether analytical or review) should be halted until the reason for the problem has been identified.  Finding the source of a QC problem involves identifying probable sources of error, and checking each source to determine if the protocols were properly followed.  Common sources of error and expected follow-up protocols are outlined on Table 11-2.  Usually, the individual who is responsible for identifying the problem is responsible for determining the cause.  However, other personnel and organizations may need to cooperate.

 

11.4            INITIATING CORRECTIVE ACTION

 

      When the source of a QC error has been identified, appropriate steps must be taken to eliminate or minimize recurrences.

 

11.4.1.  If a QC measure listed in 11.2.1 above is not acceptable, testing cannot continue until the QC check meets specifications.  Corrective actions may be initiated:

      1.   By the individual who is operating the instrument; or

      2.   By an individual in oversight authority (i.e. supervisor or QA Officer) if a solution is not immediately apparent.

 

11.4.2.      Corrective actions for QC measures in 11.2.2 and 11.2.3 must be initiated by the individual who identifies the problem.

 

11.5            SPECIFIC CORRECTIVE ACTIONS

 

      A list of expected corrective actions for each QC measure is included on Table 11-2.  Since many QC problems have unique solutions, the corrective action protocols are not limited to those listed.  Further assessment, based on an individual's experience and knowledge may be warranted.

 

11.6            DOCUMENTATION AND NOTIFICATION OF AFFECTED PARTIES

 

      If a quality control measures fails to meet acceptance criteria, the QC measure, and the procedures were used to correct the problem must be documented.

 

      Documentation does not imply a formal memo or corrective action form:

      1.      Corrective actions that are initiated during an on-going analytical run may be documented on the chromatogram, integrator or strip chart recorder records as well as in the instrument, analytical and/or field logs.  

      2.      Corrective actions that require input or intervention of more than one individual must at a minimum be documented in the related logs and records.  Corrective action forms for larger organizations are recommended.

      3.   If more than one organization is involved with identifying a QC problem and the associated corrective actions, formal memos are recommended, although dated and signed phone logs are acceptable.  In all cases, a copy of all documentation should be maintained in the project files.

 

      If an identified quality control problem affects more than one set of data or multiple projects, the documentation associated with identifying and resolving the problem must be cross referenced to all associated projects.

 

11.7            CORRECTIVE ACTIONS FROM EXTERNAL SOURCES

 

      The need to initiate corrective action may be the result of  activities or audits from external sources.  Sources include systems audits; performance audits; split samples; blind QC samples; and findings from project or data validation review.

 

      IN ALL CASES, DER RECOMMENDED CORRECTIVE ACTIONS MUST BE INITIATED.


Table 11.1

ACCEPTANCE CRITERIA AND CORRECTIVE ACTIONS FOR

QUALITY CONTROL CHECKS

 

QC CHECK

 

ACCEPTANCE CRITERIA

 

 

 

 

 

BLANKS

 

 

 

 

Method blank                                                                  

<MDL

 

 

Reagent blank  

 

 

 

Calibration blank

 

 

 

Precleaned Equipment Blanks

 

 

 

Field Cleaned Equipment Blanks

 

 

 

Trip Blanks

 

 

 

 

 

 

CALIBRATION

 

 

 

 

Initial calibration

a.  Method acceptance criteria*

 

 

 

b.  DER SOP acceptance criteria

 

 

 

 

 

 

QC Check Standard

a.  Method acceptance criteria*

 

 

 

b.  +/- 10% of true value

 

 

 

c.  Internally generated control limits

 

 

 

 

 

 

Continuing calibration

a.  Method acceptance criteria calibration*

 

 

 

b.  +/- 20% deviation from initial calibration

 

 

 

c.  Internally generated control limits

 

 

 

 

 

 

Interference standard

Method acceptance criteria

 

 

 

 

 

 

Tuning criteria

Method acceptance criteria

 

 

 

 

 

SYSTEM PERFORMANCE CHECKS

 

 

 

Pesticide

Method acceptance criteria

 

 

 

 

 

 

Standard Reference Materials

Within certified limits

 

 

 

 

 

 

QC Check Samples

Within specified limits

 

 

 

 

 

SPIKES

 

 

 

 

Matrix Spike

Within range specified by CompQAP (accuracy acceptance limits)

 

 

 

 

 

 

Blank Spike

Within method specified criteria

 


Table 11.1, cont.

ACCEPTANCE CRITERIA AND CORRECTIVE ACTIONS FOR

QUALITY CONTROL CHECKS

 

QC CHECK

 

ACCEPTANCE CRITERIA

DUPLICATES

 

 

 

Laboratory Duplicates

Within range specified by CompQAP(precision acceptance limits)

 

 

 

 

Matrix Spike Duplicates

Within range specified by CompQAP (precision acceptance limits)

 

 

 

 

Field Duplicates

Within range specified by CompQAP

 

 

(precision acceptance limits)

 

 

 

OTHERS

 

 

 

Surrogate Standards

Method acceptance criteria

 

 

 

 

Internal Standards

Method acceptance criteria

 

 

 

 

Split Samples

Meets precision criteria of CompQAP

 

 

 

 

Titrating Solutions

a.  +/- 10% of expected (lab determined) value

 

 

b.  Replicate sample aliquot results are within method specified limits

 

 

 

MICROBIOLOGY

 

 

 

Monthly parameters

Per Page 305 of EPA-600/8-78-017

 

  Chlorine Residual

 

 

  Conductivity

 

 

  Heterotrophic Count

 

 

 

 

 

Annual metals concentration

Per Page 305 of EPA-600/8-78-017

 

 

 

 

Distilled water suitability test

Per Table IV-A-3 of EPA-600/8-78-017

 

 

 

 

Incubators

35 +/-0.5 C or 44.5 +/-0.2 C

 

 

 

 

Duplicates

Within calculated precision criteria

 

 

 

 

Morphological and Biochemical Confirmation

 Per Table IV-A-5 of EPA-600/8-78-017

 

 

 

 

Positive and negative media controls

Per Table IV-A-4 of EPA-600/8-78-017

 

 

 

 

MF Blanks

<1

 

 

 

 

MPN dilution blanks

<1

 

 

 

 

Inhibitory residue

Must meet method criteria

 

 

 


Table 11.1, cont.

ACCEPTANCE CRITERIA AND CORRECTIVE ACTIONS FOR

QUALITY CONTROL CHECKS

 

QC CHECK

 

ACCEPTANCE CRITERIA

MICROBIOLOGY, cont.

 

 

 

Membrane Filter Analysis

Verified colonies must be coliforms

 

 

 

 

MPN Analysis

Verified colonies must be coliforms

 

 

 

BIOASSAY

 

 

 

One set of dilution water controls

Method acceptance criteria

 

 

 

 

Reference toxicant

a.  Per specified method criteria

 

 

b.  Outside historically derived control limits

 

 

 

SPECIES IDENTIFICATION

 

 

 

Confirmation

98-99% of unidentified species are

 

 

verified by independent expert

 


Table 11.2

PROBABLE SOURCES AND EXPECTED CORRECTIVE ACTIONS

 

1.   BLANKS

a.   Sources and expected review procedures:

      1.            Contaminated reagents - verify reagent sources

      2.            Environmental Contamination (all sample collection, sample and analysis conditions)  - review sampling handling protocols

      3.            Improper or incomplete laboratory and/or field decontamination/cleaning procedures - review cleaning protocols

      4.            Contaminated sample containers - verify source and storage conditions

      5.            Contaminated source water - verify water source

b.      Expected Corrective Actions:

      1.            Review data with respect to reported contamination levels.  If sample concentrations are near the reported blanks levels, reprocess (reextract or digest) associated samples or resample.  If sample concentrations or the reporting levels are significantly higher than blanks, or contaminants are not detected in the  samples, report the sample data and concentrations in blank.

      2.            Take measures to eliminate future problems:  discard reagents, revise protocols, perform preventative maintenance on system, adjust use of interfering chemicals (solvents, fuels, etc.).

 

2.       CALIBRATION

a.   Sources and expected review procedures:

      1.            Improperly prepared or outdated standards - review preparation logs for calculation/dilution errors and use of expired sources.

      2.            Improperly prepared or outdated check standard - verify check standard

      3.            Poor instrument response - determine if preventative maintenance is required

      4.            Incorrect calculations - review and verify all calculations

      5.            Contamination problems (see blanks above)

b.      Expected Corrective actions:

      1.            Recalculate calibration curve

      2.            Prepare fresh standards

      3.            Recalibrate instrument

      4.            Perform preventative maintenance

      5.            Perform mass calibration and retune

      6.            Reanalyze all samples bracketing those from previous ACCEPTABLE QC check through next acceptable QC check.

      7.            Take measures to eliminate sources of contamination

 

3.   SYSTEM PERFORMANCE CHECKS

a.   Sources and expected review procedures:

      1.            Pesticides:

                  Poor column performance - replace/repack column

      2.            Standard Reference Materials and QC Check Samples:

a.      Improper sample preparation or analysis - review all protocols associated with sample preparation and analysis

b.      Incorrect dilutions or calculations - recheck all calculations

c.      Contamination (see blanks above)

b.      Expected Corrective Actions:

      1.            Reanalyze all samples bracketing those from previous ACCEPTABLE QC check through next acceptable QC check

      2.            Reprocess all samples associated with QC check sample or standard reference material (unless the problem is unique to processing of the check sample)

      3.            Take measures to eliminate sources of contamination

 

4.   SPIKES

a.   Sources and expected review procedures:

      1.            Error in calculation - review/recheck all calculations

      2.            Error in preparing or using spike solutions - review all preparation and/or analytical logs (including sample preparation) for proper dilutions, solvents, buffers, etc.

      3.            Outdated standards - review expiration dates and standard preparation logs

      4.            Contamination problems (see blanks above)

      5.            Poor instrument response - determine if preventative maintenance is required

b.      Expected Corrective Actions:

      1.            Take measures to eliminate contamination problems, reprocess if necessary

      2.            Perform required maintenance and revise pm schedules

      3.            Review preparation, calculation and record keeping to determine if additional training or more stringent protocols are needed

      4.            If the laboratory has no historical data to show that the sample matrix produces consistently unacceptable (out of control) recoveries, and none of the sources discussed above are responsible for the problem, the sample must be reprocessed and reanalyzed.  If reanalysis produces the same result, associated samples should be reported with qualified results.  If results are different, all associated samples must be reprocessed for analysis.

 

5.      DUPLICATES

a.   Sources and expected review procedures:

      1.            Non representative sample - review sample collection and/or sample processing protocols

      2.            Error in calculations - recheck calculations

      3.            Contamination problems (see blanks above)

      4.            See matrix spikes above

b.      Expected Corrective Actions:

      1.            Report data with qualifiers and explanation

      2.            Revise sample collection/sample processing protocols to assure a representative sample

      3.            Takes measures to eliminate contamination problems.

      4.            Reprocess and reanalyze sample set (if laboratory generated replicate).

 

6.      SURROGATE SPIKES

a.   Sources and expected review procedures:

      1.      See 4.a above

b.      Expected Corrective Actions:

      1.      See 4.b above

 

7.   INTERNAL STANDARDS

a.   Sources and expected review procedures:

      1.      See 4.a above

b.      Expected Corrective Actions:

      1.      See 4.b above

      2.            Reanalyze samples from last acceptable QC check to next acceptable QC check

 

9.   SPLIT SAMPLES

a.   Sources and expected review procedures:

      1.      See 5.a above

b.      Expected Corrective Actions:

      1.      See 5.b above

 

10.  TITRATING SOLUTIONS

a.   Sources and expected review procedures:

      1.            Error in calculation - review/recheck all calculations

      2.            Error in preparing or using titrant and standard solutions - review all preparation and/or analytical logs (including sample preparation) for proper dilutions, solvents, buffers, etc.

      3.            Outdated standards and/or - review expiration dates and standard preparation logs

      4.            Contamination problems (see blanks above)

      5.            Non representative sample - review sample collection and/or sample processing protocols

      6.            Indistinct or inconsistent endpoint readings

b.      Expected Corrective Actions:

      1.            Take measures to eliminate contamination problems, reprocess if necessary

      2.            Review preparation, calculation and record keeping to determine if additional training or more stringent protocols are needed

      3.            If replicate analyses are not acceptable, titrate additional aliquots

      4.            Reanalyze samples from last acceptable QC check to next acceptable QC check

      5.            Train analysts to titrate to consistent endpoint

 

11.      MICROBIOLOGY - MONTHLY PARAMETERS, METALS, WATER SUITABILITY

a.   Sources and expected review procedures:

      1.      Deionizer/Water not functioning properly

b.      Expected Corrective Actions:

      1.            Clean, replace cartridges and/or perform other preventative maintenance tasks

      2.            Reanalyze water

      3.            Reprocess samples (if still within holding times) or resample

 

12.      MICROBIOLOGY - AUTOCLAVE

a.   Sources and expected review procedures:

      1.      Autoclave not functioning properly

b.      Expected Corrective Actions:

      1.      Perform preventative maintenance and resterilize

 

13.      MICROBIOLOGY - INCUBATORS

a.   Sources and expected review procedures:

      1.      Incubator not functioning properly

      2.      Thermometers or recording devices not functioning properly

b.      Expected Corrective Actions:

      1.      Perform preventative maintenance on devices, recalibrate if necessary

      2.      Reprocess samples (if still within holding times) or resample

   

14.      MICROBIOLOGY - DUPLICATES

a.   Sources and expected review procedures:

      1.            Counting errors or difficulties in identifying coliform organisms (membrane filter).

      2.            Nonrepresentative sample.

      3.            Contamination problems.

b.      Expected Corrective Actions:

      1.            Recount or re-examination colonies to determine counting error or misidentifications.

      2.            Examine blanks and samples analyzed to determine possible sources of contamination.

      3.            If 14.a.2 above is found to be the problem, or the problem has not been identified, data must be invalidated and resampling and retesting must occur.

 

15.      MICROBIOLOGY - ATCC AND BIOCHEMICAL CONFIRMATION

a.   Sources and expected review procedures:

      1.      Media prepared improperly (incorrect pH, sterilized too long, etc.).

      2.      Incorrect incubator temperatures.

      3.      Media shelf life has expired and no longer functions properly.

b.      Expected Corrective Actions:

      1.            Isolate problem with media preparation.

      2.            Confirm proper incubator temperatures.

      3.            Prepare new batch of media from the same lot to determine media acceptability or discard media if the shelf life has expired.

      4.            Prepare new media from a different lot number.

      5.            Invalidate all affected data linked to the media that was not functioning acceptability.

 

16.      MICROBIOLOGY - POSITIVE AND NEGATIVE MEDIA CONTROLS

a.   Sources and expected review procedures:

      1.      See 15.a.1-3 under ATCC or biochemical confirmation.

      2.      Samples used were not positive or negative.

      3.      Improper analytical protocol.

      4.      Contamination problems (negative control).

      5.      Colony misidentification.

      6.      Stressed organisms that did not respond in a typical fashion.

b.      Expected Corrective Actions:

      1.            See 15.b.1-4 under corrective actions for ATCC or biochemical confirmation.

      2.            Re-examine response for misidentifications.

      3.            Use alternate positive and negative control samples to confirm media response and check on original samples.

      4.            Invalidate data and retest if problem was with the media or testing system.

 

17.      MICROBIOLOGY - MF BLANKS

a.   Sources and expected review procedures:

      1.      Equipment or rinse water/dilution water improperly sterilized.

      2.      Rinsing technique not adequate.

      3.            Contamination problems.

b.      Expected Corrective Actions:

      1.            Review sterility checks on the autoclave for rinse/dilution water and other associated equipment.

      2.            Evaluate rinsing protocols between samples.

            3.            Review testing procedures and test location for other sources of contamination.

      4.            Reject data and resample

 

18.      MICROBIOLOGY - MPN DILUTION BLANKS

a.   Sources and expected review procedures:

      1.      Equipment or dilution water not properly sterilized.

      2.      Contamination problems during test procedure.

b.      Expected Corrective Actions:

      1.            Review sterility checks on the autoclave for dilution water and other sterilized equipment.

      2.            Review testing procedures and location for possible sources of contamination.

 

19.      MICROBIOLOGY - INHIBITORY RESIDUE TESTS

a.   Sources and expected review procedures:

      1.      Detergent residues inhibit bacterial growth.

      2.      Alternate rinsing practice alleviates the problem.

b.      Expected Corrective Actions:

      1.            Implement rinsing protocols that produce an acceptable inhibitory residue test.

      2.            Change detergents to one that produces an acceptable test result under normal rinsing operations.

 

20.      MICROBIOLOGY - MEMBRANE FILTER VERIFICATION

a.   Sources and expected review procedures:

      1.      Compare original counts against verified colony counts.

b.      Expected Corrective Actions:

      1.            Adjust initial colony count based upon positive verification percentage and report as verified coliform count.

 

21.      MICROBIOLOGY - MPN COMPLETED TESTS

a.   Sources and expected review procedures:

      1.      Compare original results against completed test results.

b.      Expected Corrective Actions:

      1.      Adjust original MPN result calculated from the completed test results.

 

22.  BIOASSAY - DILUTION WATER CONTROLS

a.   Sources and expected review procedures:

      1.      See 1.a.1-5 above

b.      Expected Corrective Actions:

      1.      Invalidate data and retest

 

23.  BIOASSAY - REFERENCE TOXICANTS

a.   Sources and expected review procedures:

      1.      Inappropriate test organisms

b.      Expected Corrective Actions:

      1.      Use new test organism or report data with qualifiers

 

24.  SPECIES IDENTIFICATION

a.   Sources and expected review procedures:

      1.      Outdated identification keys

      2.      Insufficient training in identification techniques 

b.      Expected Corrective Actions:

      1.      Reidentify entire set and rerun associated data analyses

      2.      Provide additional training

      3.      Obtain updated keys

 



* Method acceptance criteria shall be followed.  If none is provided, the DER criteria (b) or internal, historically generated control limits (c) shall be used.