| DEP SOPs |
|
Site index
|
4.2
AQUEOUS
SAMPLING PROCEDURES There
are several requirements that are common to all types of surface water sampling
events and are independent of technique. Several
of these requirements are concerned with sample parameters that are inherently
difficult to sample. In addition to
the below procedures, overall care must be taken in regards to equipment
handling, container handling/storage, decontamination, and record keeping.
4.2.1.1 Sample
collection equipment and non-preserved sample containers must be rinsed with
sample water before the actual sample is taken.
Exceptions to this are: oil & grease, TRPH, microbiological, VOCs, or
any pre-preserved container.
4.2.1.2 If
protective gloves are used (see Section 4.0.2), they shall be clean, new and
disposable. These should be changed prior to moving to the next sampling
point.
4.2.1.3 Sample
containers for source (i.e. concentrated wastes) samples or samples suspected of
containing high concentrations of contaminants shall be placed in separate
plastic bags immediately after collecting, preserving, tagging, etc.
4.2.1.4 If
possible, ambient, or background samples should be collected by different field
teams. If separate collection is not possible, the ambient or
background samples shall be collected first and placed in separate ice chests or
shipping containers. Highly
contaminated samples shall never be placed in the same ice chest as
environmental samples. It is a good
practice to enclose highly contaminated samples in a plastic bag before placing
them in ice chests. Ice chests or
shipping containers with samples suspected of being highly contaminated shall be
lined with new, clean, plastic bags.
4.2.1.5 If
possible, one member of the field team should take all the notes, fill out tags,
etc., while the other member does all of the sampling.
4.2.1.6 Teflon
or glass is preferred for collecting samples where trace contaminants are of
concern. Equipment constructed of
rubber or plastic (e.g., PVC, Tygon, most Van Dorn Samplers) shall not be used
to collect samples for trace organic compound analyses. 4.2.2
Special Parameter - Specific Handling Procedures
1. Since the
concentration standards and/or guidance criteria for many analytes are in the
(sub)parts per billion range, extreme care must be taken to prevent
cross-contamination.
2. Most
of the parameter groups listed in sections 4.2.2.1 through 4.2.2.8 below, shall
be taken as grab samples unless Department requirements dictate otherwise.
The exceptions are extractable organics and total metals which may be
taken as composites, if required.
3. There
is a greater chance of cross contamination when collecting composites because of
increased sample handling and more equipment.
4. The
following eight categories of parameters have specific sampling techniques and
considerations which must be followed to collect unbiased, uncontaminated
samples.
THE PROCEDURES OUTLINED BELOW SHALL BE USED FOR ALL AQUEOUS SAMPLING (I.E. SURFACE
WATER, WASTEWATER, GROUNDWATER, STORMWATER ETC.).
4.2.2.1
Metals Sampling a.
Sample containers
1.
New or properly cleaned plastic containers may be used for metals
sampling. Glass bottles may also be
used, but they are prone to breakage and occasionally react with the sample to
either leach or adsorb metals from the glass itself.
2.
All containers for metals sampling, new or previously used, shall be
cleaned by following protocols outlined in Section 4.4.1.
3.
Visually inspect polyethylene or glass containers for defects or
contamination. Discard if defects
are present or containers do not appear clean. b.
Preservation
1.
Samples shall be preserved with nitric acid (HNO3) of a grade that is
suitable for use in trace metals analysis.
2.
Preservation shall occur within 15 minutes of sample collection or
filtration (if applicable) unless collected as a 24-hour composite (see
4.2.4.6.b.5).
3.
Adequate HNO3 shall be added per liter of sample to reduce the pH to
below 2.0 to keep metals in solution and prevent them from adsorbing or
absorbing to the container wall.
4.
If only dissolved metals are to be measured, the sample shall be filtered
immediately after sample collection through a 0.45 um membrane filter.
The sample shall not be preserved before filtration. See Table 4.1 for approved filtration equipment.
5.
Samples submitted for Chromium VI should not be acidified. c.
Sample collection protocol:
1.
Remove the cap from the sample container and rinse container with sample
water (if not pre-preserved). Carefully
pour sample into the container without allowing the sampling device to touch the
rim of the sample container.
2.
If adding preservatives in the field, the sample container should not be
filled to capacity.
3.
Acidify the sample to pH of 2 or less by adding a measured quantity of
concentrated HNO3 or 1+1 HNO3 into the container.
4.
NOTE: If containers are pre-preserved by a subcontract laboratory, the
sample must be poured into the container slowly to prevent the acid from
splattering. As a precautionary
note, the addition of water to acid can generate enough heat to burn unprotected
hands.
5.
Tightly cap the sample container and shake to distribute the acid.
Pour an aliquot of the acidified sample into a disposable container (e.g.
sampling cup) or onto a piece of NARROW range pH paper to determine if the pH is
less than 2.0. DO NOT PUT THE pH
PAPER DIRECTLY INTO THE SAMPLE CONTAINER! a.
Field experience has shown that UNDER NORMAL CIRCUMSTANCES, 2 ml of
concentrated HNO3 added to 250 ml of sample water will reduce the pH to less
than 2. b.
If the pH is greater than 2, add additional
MEASURED amounts of acid until the pH has been reduced. c.
Record the total amount of acid that was added to the sample.
This documentation is necessary for the next site visit, since additional
acid may need to be added to the sample on subsequent visits. d.
Acidify at least one of the equipment blank(s) with the GREATEST amount
of acid that was required in the sample set and note the amount in field
documentation.
6.
Following proper sample preservation, tightly cap, affix a sample label,
apply a seal (if required), and complete the COC or laboratory transmittal form.
7. Aqueous
samples for metals need not be cooled to 4 C.
8.
Make a note on the transmittal form identifying samples that have
entrained sediment.
d.
Filtration
1.
For certain studies or projects, it may be necessary to obtain dissolved
(i.e. filtered) samples. All
samples that are filtered shall be identified in field notes and on final
reports as "dissolved" or "filtered" metals.
2.
Specific protocols for collecting dissolved metals from groundwater
samples are discussed in Section 4.2.5.6.g.
Filtered samples SHALL NOT be collected from groundwater sources unless: a.
The Department has required or approved the protocol; or b.
The organization feels that the contamination is directly related to the
suspended material rather than the groundwater.
In this case, BOTH unfiltered and filtered samples shall be collected and
analyzed. The results shall be
submitted to the Department for review. Based
on the data, the Department may require continued collection of unfiltered
samples, BOTH unfiltered and filtered samples or only filtered samples.
3.
Surface water samples may use the sample protocols that are specified for
groundwater (Section 4.2.5.6.g.) These
protocols are recommended when sampling static surface water sources (i.e.
subsurface samples from lakes, ponds, lagoons or ocean) since exposure to air
can change the concentration of metals in solution.
When sampling from moving sources (i.e. rivers or streams) or just below
the surface, filtered samples may be collected into an intermediate container
and filtered with syringe-type or tripod type filtration units.
4.
Allowing a sample to settle and decanting the supernate (upper water
layer) has been proposed as a means of removing suspended material.
This technique MAY NOT be used for groundwater samples, and is not
recommended for other sources because: a.
Settling times techniques are highly dependent on particle size and
concentration and may not be reproducible; b.
Preservation for metals must occur within 15 minutes of sample collection
which may not be sufficiently long for highly turbid samples to settle; and c.
The analytical results cannot be reported as "total" or
"dissolved".
If this technique is used, the following
protocols must be followed: a.
Samples shall not be acidified before settling occurs; b.
Total time for settling shall not exceed 15 minutes; c.
The resultant supernate shall be carefully decanted into an appropriate
container and preserved using protocols described above; d.
Field notes shall specify the length of time the sample was allowed to
settle, as well as observations on the initial and final (supernate); and e.
The final report shall identify the technique that was used to collect
the sample (i.e. decanted). f.
NOTE: samples SHALL NOT be transported back to the laboratory for
settling, UNLESS entire procedure (transport, settling, decanting and
preservation) can occur within 15 minutes of sample collection. 4.2.2.2
Extractable Organics and Pesticides
Conventional sampling practices shall
incorporate the following special considerations. Oil & Grease (O&G) and Total Recoverable Petroleum
Hydrocarbons (TRPH) shall follow protocols outlined in Section 4.2.2.5 below.
a.
Sample containers
1.
Collect all samples in glass containers (1 liter to 1 gal.) with
Teflon-lined caps. Note:
Teflon containers are also acceptable.
2.
Amber glass should be used for PAHs, nitrosamines, nitroaromatics, and
isophorone.
3.
Visually inspect glass bottles to assure that there are no glass or liner
defects. If defects are present
and/or the sample containers do not appear clean, the bottles must be discarded.
4.
Sample containers must be cleaned according to the protocols specified in
Section 4.4.1.
5.
Composite samples from automatic WW samplers must be collected in
refrigerated glass containers through Teflon tubing. b.
Preservation
1.
Tables 4.2 and 4.3 must be followed to determine the specific
preservation method for each group of organic compounds and pesticides.
2.
All samples must be placed on wet ice immediately after collection.
Samples must be maintained at a temperature of 4 C.
3.
If the samples for pesticides cannot be extracted within 72 hours of
collection, the sample pH must be in the range of pH 5 to 9.
If needed, sample must be adjusted to the specified pH range with sodium
hydroxide or sulfuric acid.
4.
Other extractable samples need not be pH-adjusted with acid or base.
5.
Samples must be extracted within 7 days of sample collection and the
extracts analyzed within 40 days of extraction.
6.
If residual chlorine is present, sodium thiosulfate must be added. c.
Sample collection protocol:
1.
Sample bottles should be prerinsed with sample before collection, except
Total Recoverable Petroleum Hydrocarbons (TRPH), Oil & Grease, etc. (see
4.0.3) or any prepreserved sample container.
2.
Remove the cap from the bottle without touching the Teflon liner.
3.
Do not allow the sampling equipment or hands to touch the rim of the
sample container. a.
For bailer sampling, it may be necessary to utilize a stainless steel or
Teflon delivery tube (fits into the bottom of the bailer).
4.
Fill bottle with sample to almost full capacity.
5.
Quickly place the Teflon lined cap over the bottle and tighten securely.
6.
Affix a sample label, seal (if required), and complete the
chain-of-custody form.
7. Put
the sample bottle in a plastic sample bag and place on wet ice immediately.
8.
Make a note on the lab transmittal form identifying samples that appear
highly contaminated or exhibit other abnormal characteristics (i.e. foaming,
odor, etc.). 4.2.2.3
Volatiles Sampling a.
Sample containers
1.
Analysis of volatile organic substances requires a glass sample vial,
sealed with a teflon-coated septum.
2.
AT A MINIMUM, duplicate samples must be collected, although some
laboratories require three or more vials. If
the containers are not supplied by the laboratory, verify the laboratory's
policy on how many vials are necessary and collect the specified number of
vials.
3.
Visually inspect the glass vials to assure that there are no glass or
septum defects (e.g. rim must have not nicks or visible depressions); septum
must not be deformed, etc.). If
defects are present and/or sample containers or septums do not appear to be
clean, the vials must be discarded.
4.
Sample vials may be purchased precleaned from commercial vendors, or must
be cleaned according to protocols outlined in Section 4.4.1.
5.
NOTE: VIALS FOR VOCS ARE NOT
RINSED WITH SAMPLE. b.
Preservation
1.
Table 4.2 must be followed to determine the specific preservation method
for each group of volatile organic compounds.
2.
If residual chlorine is not present, the vials shall be filled with the
sample, acidified (prepreserved containers are acceptable) with HCl and labeled
"preserved".
3.
If the volatile aromatics are to be analyzed within 7 days, HCl is not
necessary.
4.
Sodium thiosulfate must be added to samples with residual chlorine (see
sampling protocols below).
5.
Samples must be placed on wet ice immediately after sample collection.
A temperature of 4 C must be maintained until the sample has arrived at
the laboratory. c.
Sample collection protocols:
1.
All fuel or exhaust sources which could cause VOC contamination must be
situated well away and downwind of the sampling site (see Section 4.0.5). a.
If possible, fuels should be transported and stored in a separate vehicle
from empty vials and collected samples. b.
All petroleum fueled engines (including the vehicle) must be situated
downwind of the sampling operations.
2.
Samples shall not be aerated during sample collection. a.
Extreme caution must be exercised when filling a vial to avoid any
turbulence which could promote volatilization.
b.
Carefully pour the sample down the SIDE of the vial to minimize
turbulence. As a rule, it is best
to gently pour the last few drops into the vial so that surface tension holds
the water in a "convex meniscus."
3.
Do not allow the sampling equipment to touch the rim of the sample
container. a.
For bailer sampling, it may be necessary utilize a stainless steel or
Teflon delivery tube or “pigtail" to
obtain a gentle trickle of sample into the vial. b.
It is sometimes difficult to completely fill the vial directly from some
waste streams. The sample may be
collected in a precleaned intermediate sample collection device made of the
appropriate materials (see Table 4.1) and carefully poured into the VOC vials.
4.
The investigator must determine if the water to be sampled contains
residual chlorine. a.
If residual chlorine is present; add 10 mg of sodium thiosulfate to the
vial (laboratory may supply vials with premeasured quantities). b.
Fill the vial 90% with sample. c.
Add four drops of concentrated HCl (more acid may be needed if the sample
is known to contain high levels of bicarbonate or is otherwise buffered).
Add additional sample (if needed) to create a convex meniscus and cap
with zero headspace (see 5 below). d.
Label vial appropriately (preserved/sodium thiosulfate/acid).
5.
The sample must be collected so that there are no air bubbles in the
container after the screw cap and septum seal are applied. a.
Vial must be filled so that the sample surface is above the container rim
(convex meniscus). b.
The cap with the septum is then quickly applied (make sure teflon side of
septum is down). Some sample may
overflow, but air space in the bottle must be eliminated.
c.
If acid has been added to the sample, tip the vial gently two or three
times to distribute the preservative. d.
Turn the bottle over and tap it to check for bubbles.
1.
If any are present, remove the cap, add a few more drops of sample, recap
and test for bubbles. REPEAT NO
MORE THAN 3 TIMES.
6.
Sampling and preservation containers may be prelabeled prior to any field
activities. This may reduce confusion during a sampling event.
7.
All the vials must be labeled. Make
note in the field records of any samples that appear highly contaminated or
appear to effervesce when acid is added. NOTE: If the sample reacts with the acid by generating gas, DER
recommends collecting unpreserved samples for analysis (seven-day holding time
must be met).
8.
Wrap each vial in bubble-wrap, or equivalent, and place each vial in a
small ziplock-type bag and immediately place on wet ice.
9.
Complete field records.
10.
Protect samples from environmental contamination during storage and
transport to the laboratory (4.2.2.3.c.1 above). a.
As an added measure, replicate samples may be sealed in a container with
vermiculite. This will add further
protection from potential contamination. 4.2.2.4
Bacteriological Sampling a.
Sample containers
1.
Samples must be collected in containers that have been sterilized
according to Standard Methods (17th Edition) or the EPA's Microbiological
Methods for Monitoring the Environment, 14th edition. a.
Presterilized Whirlpak bags (or equivalent) are typically used for
sampling. b.
If Whirlpaks are not used, a 125 ml or larger sample container must be
used to provide a minimum sample volume of 100 ml and adequate mixing space.
2.
Unlined caps or ground glass tops shall be used to ensure complete
sterilization of the container's closure.
3. Bottles
and caps shall be sterilized according to protocols outlined in Section 4.4.1 or
purchased presterilized from a commercial vendor. b.
Preservation
1.
Samples shall be preserved according to Tables 4.2 and 4.5.
2.
All samples shall be place on wet ice immediately after sample
collection.
3.
When sampling water containing residual chlorine, a dechlorinating agent
such as sodium thiosulfate must be added to the sample container.
a.
The final concentration of sodium thiosulfate in the sample shall be
approximately 100 milligram per liter (mg/L) in the sample. b.
As a general rule, this concentration may be achieved by adding 0.1 ml of
a 10 percent solution of sodium thiosulfate to a 125 ml sample bottle. c.
The dechlorinating agent neutralizes any residual chlorine and will
prevent further reaction between bacteria and chlorine.
4.
ANALYSIS MUST COMMENCE WITHIN 6 HOURS FOR NON-POTABLE SOURCES AND 30
HOURS FOR POTABLE SOURCES. Special laboratory arrangements may need to be made so that
the holding times are not compromised (may require local lab analysis). c.
Sample Collection Protocols:
1.
Bacteriological sampling must always be collected as a grab sample and
must never be composited.
2. The
container must be kept unopened until the moment that the sample is collected.
3.
DO NOT RINSE CONTAINER BEFORE COLLECTING SAMPLE.
4.
When the Whirlpak bag or sample bottle must be lowered into the waste
stream, either because of safety or impracticality (manhole, slippery effluent
area, etc.), care must be taken to avoid contamination.
5.
Samples shall never be collected in an unsterilized sample container and
transferred to a sterile container.
6.
Be careful not to put fingers into the mouth of the container or on the
interior of the cap.
7.
If sampling intermediate sampling devices (i.e. bailers) or from in-place
plumbing, the sampling device or the tap do not need to be disinfected (i.e.
swabbing with alcohol or flaming with heat source). a.
Intermediate sampling devices shall be thoroughly rinsed with sample
water prior to collecting the sample. For
this reason, microbiological samples should be among the final samples that are
collected from the sampling location. b.
Spigots shall be flushed at maximum velocity (see Sections 4.2.6 and
4.2.7) to purge the system and remove particulates.
Sample flow shall be reduced to approximately 500 ml/min and allowed to
run a few minutes before collecting samples (or microbiological samples may be
collected last). DO NOT STOP FLOW
BEFORE OR DURING THE FILLING PROCESS.
8.
Surface water sample collection: a.
To sample with a rigid container, hold the bottle near the base and
plunge neck downward, below the surface. Turn
container until the neck points slightly upward with the mouth directed toward
the current. Fill to within about
1/2 inch of the top and recap immediately. b.
To sample with a Whirlpak bag, open the bag by zipping off the top and
pulling the white tabs to open the bag. Hold
the bag in the hand or attach to a long handle and plunge neck downward and up
in one sweeping arc.
9.
Intermediate containers (i.e. bailers) a.
Obtain sufficient sample in the sample collection device to completely
fill the sample container. b.
Begin pouring sample out of the device BEFORE collecting into the
container. c.
Continue to pour sample out of the device, place container under flowing
stream, and fill. DO NOT STOP FLOW
BEFORE OR DURING THE FILLING PROCESS.
10.
Wells with in-place plumbing, spigots and/or faucets a.
Samples shall be collected after flow has been reduced to 500/ml per
minute. b.
Allow the water to flow at the reduced rate for a few minutes before
collecting the sample, or collect all other samples prior to taking the
microbiological sample. c.
DO NOT STOP FLOW BEFORE OR DURING THE FILLING PROCESS.
4.2.2.5 Oil
and Grease (O&G) and Total Recoverable Petroleum Hydrocarbon
(TRPH) Sampling a.
Sample Containers
1.
Samples for O&G and TRPH shall be collected in 1 liter wide-mouth
glass bottles. a.
The lid shall be teflon-lined. b.
If the cap is not teflon-lined, a sheet of teflon extending out from the
lid may be used.
2.
Visually inspect glass bottles to assure there are no glass or cap
defects. If defects are present
and/or sample containers do not appear to be clean, the bottles should be
discarded. b.
Sample Preservation
1.
Since losses of the product will occur on sampling equipment, composite
samples shall not be collected.
2.
The sample must be immediately preserved by adding H2SO4 or HCl to reduce
the pH to 2.0 or less.
3.
Samples must be placed on wet ice immediately after preservation.
The temperature of the sample must be maintained at 4 C until received
and processed by the laboratory. c.
Selection of Sampling points
1.
Oil and grease may be present in wastewater as a surface film, an
emulsion, a solution, or as a combination of these forms.
Since it is very difficult to collect a representative ambient sample for
oil and grease analysis, the sampler must carefully evaluate the location of the
sampling point.
2.
The most desirable sampling location for both O&G and TRPH is the
point where greatest mixing is occurring. Quiescent
areas should be avoided, if possible.
3.
Skimming the surface for the sample is unacceptable.
4.
For compliance samples at a facility you may want to take samples at the
worst place.
5.
Neither the container, nor the sampling device, shall be rinsed before
the actual sample is taken.
6.
COMPOSITE SAMPLES SHALL NOT BE COLLECTED.
If composite data is required, individual grab samples that are collected
at prescribed time intervals must be analyzed separately to obtain the average
concentrations over an extended period. d.
Sampling Protocols
1.
Sampling for these products is unique because they are immiscible and
tend to adhere to the sampling device; therefore, these sample shall always be a
grab sample.
2.
The sample, when collected, should not be transferred to another
container. The analytical methods
require the use of the entire sample. In
addition, the sample container must be rinsed with solvent as a part of the
laboratory analytical process. Therefore
these samples must be separate and discrete samples that will be used only for
the O&G or TRPH analysis.
3.
Remove the cap from the glass bottle without contacting the interior of
the container or lid.
4. DO
NOT RINSE THE BOTTLE WITH SAMPLE WATER.
5.
Whenever possible samples should be collected directly into an
unpreserved sample container. If
intermediate sampling equipment is used, do not allow the sampling equipment to
touch the rim of the sample container. AUTOMATIC
SAMPLERS SHALL NOT BE USED TO COLLECT THESE TYPES OF SAMPLES.
6.
Fill the bottle with the sample water to almost full capacity.
7.
Add preservatives and check the pH using the protocols outlined in
4.4.2.2.b.
8.
Quickly cap the container and tighten securely.
9.
Affix a sample label, seal (if required), and complete the
chain-of-custody form.
10. Protect
glass container from breakage ("bubble wrap" is recommended), place
the sample bottle in a plastic sample bag and keep it cool to 4 C on wet ice.
11. Make
a note on the lab transmittal form identifying samples that may be highly
contaminated or any other unusual observations. 4.2.2.6
Radiological Sampling (Excludes Radon) a.
Sample containers
1.
Polyethylene, polyvinyl chloride (PVC), or Teflon containers are
recommended for collecting radioactive samples because these containers are less
adsorbent than glass or metal containers. Since
radioactive elements are often present in extremely low quantities, a large
fraction of the elements may be lost by adsorption on containers or glassware
surfaces used in analyses. This
loss may, in turn, cause a loss of radioactivity and possibly contaminate
subsequent samples due to reuse of inadequately cleaned containers.
Glass bottles are also more susceptible to breakage during handling than
plastic containers.
2.
Containers shall be cleaned according to the protocols specified in
Section 4.4.1. b.
Preservation
1.
The preservation technique for radiological sampling is acidification to
a pH of less than 2.0 with concentrated or 1+1 nitric acid (HNO3).
2. The
pH shall be checked in the field following the protocols described 4.4.2.2.b. c.
Sampling Protocols
Prior to sampling, the area may be surveyed
with a beta-gamma survey instrument, such as a Geiger-Maller meter.
If radiation levels are above instrument background, the investigator
should consult a radiation safety specialist to determine appropriate safety
procedures.
4.2.2.7
Radon Sampling Radon
is a gas and is easily removed from water sources.
Therefore, the same precautions and care used to collect volatile organic
samples shall be followed. It is
extremely important to minimize contact with air during sample collection. a.
Sample Containers
1.
Glass sample vials shall be obtained from the analyzing laboratory and
shall contain a premeasured portion of the scintillation "cocktail".
2.
A minimum of two samples is required.
Laboratories are expected to provide the sampler with the requisite
number of containers for each sample. b.
Preservation - the scintillation cocktail is the only required
preservative. c.
Sampling Protocol
The laboratory should provide specific sample
collection instructions that must be followed.
These protocols should included proper handling as well a sample size and
packing instructions. The following
are general instructions that should be used:
1.
Carefully fill a syringe (usually 10 ml) with sample water so that air
bubbles are not pulled in with the sample before, during or after filling.
2.
Place the tip of the syringe BELOW the scintillation cocktail and slowly
dispense the sample BENEATH the cocktail surface.
3.
Replace lid and cap tightly.
4.
Generally the vial is used in the laboratory analytical instrument and
labels or ID numbers on the sides of the containers may interfere with the
analysis. Check with the laboratory
for proper placement of labels or field ID numbers.
5.
Ship in an upright position in the shipping containers that have been
provided by the laboratory. If none
are provided, protect vials from breakage (bubble wrap is recommended),
segregate replicate samples in separate plastic bags, and ship to laboratory in
an upright position. 4.2.2.8
Cyanide Sampling Cyanide
is a very reactive and unstable compound. Cyanide
should be analyzed as soon as possible after collection.
Note that the currently approved method (Standard Methods and EPA MCAWW)
is being rewritten due to some inaccuracies in the method language. a.
Sample Containers
1.
The sample container shall be polyethylene or glass.
2. Containers
shall be cleaned in accordance with protocols outlined in Section 4.4.1 of this
manual. b.
Preservation
Sulfides tend to be a regional problem in
Florida and can cause interference with cyanide analyses.
1.
Proper preservation of aqueous cyanide samples must follow the standard
procedures listed below.
2.
All samples shall be tested for sulfides with test papers (EM,
Chemometrics) or kits (HACH). However, these tests may not detect sulfides in
low enough concentrations to be useful.
3.
Ultimately, all samples shall be preserved to a pH of greater than 12
with sodium hydroxide and placed on wet ice immediately after preservation.
A temperature of 4 C shall be maintained until analysis begins at the
laboratory. The pH of the samples
shall be checked to assure proper pH (see 4.4.2.2.b).
4.
Samples that may or may not contain sulfides must be preserved in one of
two ways: (1) samples are tested for sulfides, preserved with NaOH to a pH>12
and sent to the lab for analysis within 24 hours; or (2) tested for sulfides and
pretreated as follows: a.
Test for sulfides
1)
Samples with visible particulates must be filtered. Keep this filter
(#1);
2)
Sulfides may be tested with test papers (EM, Chemometrics) or kits
(HACH). Tests may not detect sulfides in low enough concentrations to be useful;
3)
Remove sulfide by adding cadmium (or lead) nitrate (or carbonate) powder
to the sample (filtrate) to precipitate the sulfides;
4)
Test for presence of sulfides. Repeat steps 2 and 3 until the test shows
no sulfides are present;
5)
Remove the precipitate (sulfides) from the sample by filtration and
discard this filter. b.
Preservation
1)
Reconstitute the sample by adding the solids collected on filter #1 back
into the filtrate;
2)
Add NaOH until the sample pH > 12 and cool to 4 C;
3)
Maximum holding time is now 14 days;
4)
Equipment blanks must be handled the same as real samples.
5.
All samples known to contain oxidizing agents (chlorine) must first be
tested as follows: a.
Test sample with KI-starch paper; b.
Add a few crystals of ascorbic acid, mix sample and retest. c.
Continue to add ascorbic acid until the test is negative; d.
Add an additional 0.6 grams of ascorbic acid per liter of sample to
remove chlorine.
4.2.3.1 Introduction
and Scope [[Surface water samples may be taken
for several reasons. Ambient
conditions can be documented for a single point in time to determine if
threshold Water Quality parameter limits are being met.
Long-term sampling of a site or watershed may be used to document trend
analysis. All of these items can be
used to determine if a particular surface water body (or segment) is meeting its
designated use. Sediment sampling
may provide confirmatory information as to contaminants that are presently contained
in the water column as well as an historical account of those contaminants that
settled or precipitated out of the water column.
Physical conditions of the water body and physico-chemical properties of
the contaminants will determine their movement in and out of the water column.
Additional information on data use, study objectives, etc. can be found
in the EPA Region IV SOP & QAM, February 1991, Section 4.8.1.
This document will henceforth be referred to as the EPA SOP. Sample site selection is dependent upon
the three major groupings; lakes, estuaries, and streams.
Sites may have already been predetermined by a EPA, DER, WMD, etc. permit
or by designation as a permanent monitoring station (PMS).
If these sites have not already been designated as above or by the DER
water or waste program for which this sampling event has been required, then
refer to the general descriptions in the EPA SOP, Sections 4.8.2.
The EPA document provides generalized descriptions for proper site
choices. Only sampling procedures
will be described in the following sections.]] This section presents the standard
operating procedures that shall be employed during field investigations to
ensure that representative surface water samples are collected.
The particular surface water types that will be addressed include; static
lakes, ponds, and impoundments; tidally-influenced estuarine areas; as well as
streams and rivers. The importance
of proper sampling of this varied matrix cannot be overemphasized.
Proper sampling for the submerged (or emergent) sediments that underlie
these surface water bodies is equally important.
Care should be taken so that samples are neither altered nor contaminated
by sample handling procedures. This section discusses grab,
depth-specific, and depth composited surface water samples.
Information regarding flow- or time-weighted aqueous sampling is found in
the Wastewater Sampling section.
4.2.3.2 General Access
will be left up to the sampling group. Ease
of access should not be the main criteria for sampling site choice.
If sampling from a bridge, by boat, or by wading, there are certain
precautions that must be considered: a.
If sampling with a boat, samples should be taken from the bow, away and
upwind from any gasoline outboard engine (see 4.0.5.1). b.
Collect samples upstream from the body when wading in to collect water
samples. c.
Care should be taken not to disturb sediments when taking samples in
lakes, ponds, impoundments. d.
If water samples and sediment samples are to be taken from the same area,
the water samples must be taken first. e.
Sampling at or near structures (dams, weirs, bridges) may not provide
representative data because of unnatural flow patterns. f.
Surface water and/or sediments should always be collected from downstream
to upstream.
4.2.3.3 Sample
Acquisition Three
(3) types of general sample acquisition methods will be discussed:
grab samplers; mid-depth samplers; and composite samplers. a.
Grab Sampling
1.
If the sample media is homogenous, grab samples are an effective and
simple technique. If homogeneity is
not known (and should never be assumed) then other techniques must be used.
2.
Surface grabs using unpreserved sample containers are encouraged since
the sample container is used for collecting the sample and, after appropriate
preservation, the same container can be submitted for laboratory analysis.
This reduces sample handling and eliminates potential contamination from
other sources (i.e. additional sampling equipment, environment, etc.). If the laboratory provides prepreserved sample containers,
the sample shall be collected in an UNPRESERVED sample container or with
sampling equipment. The container
or equipment shall be of appropriate construction (see Table 4.1) and the sample
shall be transferred immediately into the prepreserved sample container
3.
Simple Grab Samples - Typical sample collection equipment includes not
only sample containers, but also precleaned beakers, buckets, and dippers.
These samplers must be constructed appropriately (including handles): a.
Sample Container (unpreserved)
1.
submerge the container, neck first into the water,
2.
invert the bottle so the neck is upright and pointing into the water flow
(if applicable),
3.
return the filled container quickly to the surface,
4.
shake to rinse the interior surface of the container and pour contents
out downstream of sample location (see restrictions outlined in 4.0.3)
5.
Collect sample as described in steps 1,2 and 3 above.
6.
pour out a few mls of sample downstream of sample collection.
This allows for addition of preservatives and sample expansion
7.
Securely cap container, and label. b.
Intermediate vessel
1.
Collect sample as outlined in 3.a above.
2.
Pour into prepreserved sample container (or field preserve per Section
4.4.2.2.a), check pH per Section 4.4.2.2.b (if applicable), cap, and label.
4.
Pond Sampler - Another effective technique is using a pole-mounted flask,
beaker, or container. A long,
telescoping pole (swimming pool supply) is outfitted with a (non-contaminating)
clamp. An appropriately constructed
and shaped container is fitted into the clamp.
In this way the sample can be taken away from the shore, boat, bridge,
etc. and at a specific spot. The
sampling vessel can be constructed of all-inert material so that all parameters
can be sampled. a.
Submerge the clamped container neck first, invert and withdraw from
water. b.
Be careful not to entrain sediments or skim the water surface. c.
Rinse container (restrictions specified in Section 4.0.3 must be
observed), resubmerge and collect sample. Retrieve
the pole, clamp, and container and fill the sample containers.
5.
Pump and Tubing - Although the use of a peristaltic pump and tubing can
provide an adequate mid-depth or depth composite, it can also be used for taking
a grab sample. This would be
especially helpful if a large amount of sample is needed. a.
Lower appropriately precleaned tubing to a depth just below the water
surface (6 - 12 inches). b.
Turn the pump on. c.
Allow several tube volumes to pump through the system to acclimate the
tubing. d.
Make sure the tubing does not come out of the water and inadvertently
pull some surface skim water through the tubing (this may bias sample results). e.
Fill the individual sample bottles via the discharge tubing.
NOTE: THIS TECHNIQUE IS NOT ACCEPTABLE FOR OIL & GREASE, TRPH
OR VOCs. It is not recommended for
extractable organics (requires the organic trap setup, see Fig. 4.1) or
microbiologicals (new, unused tubing, including tubing in the sampling head are
required at each sampling location). b.
Mid-Depth Sampling
1.
Mid-depth samples or samples taken at a specific depth can approximate
the conditions throughout the entire water column. a.
One sample may be taken when the water body is assumed to be homogenous. b.
Additional samples can be taken from different depths at one spot to get
a much more exact approximation of the conditions. c.
Many times a single site will be sampled from: just below the surface;
mid-depth; and just above the bottom (sediment).
d.
Accurate sampler location is imperative for this sampling technique.
2.
The equipment that may be used for this type of sampling are:
a device designed specifically for depth-specific sampling (kemmerer,
niskin, beta, etc.); pumps with tubing; or double check valve bailers. a.
Samplers are available from many manufacturers and in a variety of
configurations and construction materials. b.
When purchasing and choosing a device for a particular sampling event,
please be aware that certain construction material details may preclude its use
for certain parameters (see Table 4.1):
1.
Many kemmerer samplers are constructed of plastic and rubber which
precludes their use for all organic sampling parameters (volatile and
semivolatile).
2.
Some of the newer devices are constructed of stainless steel or are
all-Teflon or Teflon coated. These
would be acceptable for all parameter groups without restriction.
3.
NOTE THAT ALL RELATED COMPONENTS (STOPPERS, ETC.) MUST BE CONSTRUCTED OF
INERT MATERIAL AS WELL IF ORGANICS ARE TO BE SAMPLED.
3.
Kemmerer, niskin, and beta type devices a.
Separate and specific deployment discussions are not provided in this
document. Manufacturers suggestions
shall be followed for specific procedures.
1.
Before lowering the sampler, measure the water column to determine
maximum depth and sampling depth.
2. The
line attached to the sampler should be marked with depth increments so that the
sampling depth can be accurately recorded.
3.
When dropping the sampler to the appropriate depth, it should be done
slowly so that sediments are not stirred up.
4.
Once the desired depth is reached, send the messenger weight down to trip
the mechanism.
5.
The sampler should be lowered and retrieved slowly.
6.
The first sample shall be discarded into a bucket (to be dumped at
conclusion of sampling).
4.
Double check-valve bailers a.
Sampling with these type of bailers shall follow the same protocols
outlined in 3 above. b.
Although not designed specifically for this kind of sampling, it will be
acceptable when a mid-depth sample is required. c.
Note: this sampler does not perform as well as the devices
described above or the pump and tubing described in the next section. d.
As the bailer is dropped through the water column, water will be
displaced through the body of the bailer. The
degree of displacement is dependent upon the check valve ball getting out of the
way and allowing water to flow freely through the bailer body. e.
An open-top bailer may also be used, but is not recommended. The
open-top arrangement will not prevent water from being exchanged in the top
portion of the bailer. f.
A closed-top bailer does not allow free water displacement on descent at
all and is not acceptable. g.
The bailer should be dropped slowly to the appropriate depth.
Upon retrieval, the (two) check valves seat, preventing water from
escaping out of or entering the bailer.
5.
Pump and Tubing a.
The most portable pump for this technique is a (12 volt) peristaltic
pump. b.
Appropriately precleaned silastic is required in the pump head and HDPE,
Tygon, etc. tubing is attached to the pump. c.
Measure the water column to determine the maximum depth and the sampling
depth. d.
Tubing will need to be tied to a stiff pole or be weighted down so the
tubing placement will be secure.
1. A
lead weight should not be used.
2.
Any dense, non-contaminating, non-interfering material will work (brick,
SS weight, etc.).
3.
Tie the weight with a lanyard (braided or monofilament nylon, etc.) so
that it is located below the inlet of the tubing. e.
Turn the pump on and allow several tubing volumes of water to be
discharged before taking the first sample. f.
Sample containers are then filled in the proper order, preserved,
labeled, and placed on ice (if required). c.
Composite Sampling
Composite sampling will be used when a single
sample that approximates a given depth interval is desired.
Any of the devices described in mid-depth sampling can be used for
composite sampling. The devices
must be activated or manipulated in a way that the actual volumes sampled within
the interval are ALL EQUAL PROPORTIONS.
For instance, because of head pressure, the pump and tubing will pull a
greater volume of sample at 5 feet in comparison to 20 feet.
For this reason, great care must be used so that sample results are not
biased. The use of the niskin, kemmerer, beta, bailers, etc.
containers may take more time, but sample control will be greater. 4.2.4.1
Introduction and Scope
Prior to mobilizing, the sampler must decide what kind of samples to
collect, for what parameters, and where to collect them.
This section will provide the guidelines as to what kind of sample should
be collected, where it should be collected, and how to collect it.
It will also discuss choosing parameters for analysis. Care must be taken to ensure that the sampling location is
correct and that the samples are representative of the discharge.
The site must be consistent with its permit.
This section is also applicable to stormwater runoff sampling. [[4.2.4.2
Site Selection
The following discussion deals with site
selection, sampling points and sample collection strategies and are for
educational and informational use. a.
Samples shall be collected at the appropriate permitted locations and at
locations necessary to determine environmental impact (e.g., effluent outfall,
ground water monitoring well, land application site, receiving water stations).
1.
If the permitted sampling point is not adequate for collecting a
representative sample, the sampler should determine the most representative
sampling point available and collect samples at both locations.
This should be done with the concurrence of DER and the facility.
2.
The reason should be documented in the field log for later resolution if
challenged and for consideration of correcting the sampling point during permit
renewal.
3.
Recommendations for a change in sampling location should be given to the
DER permit writer. Sample locations should be specified in such detail that
anyone could follow the directions to the site and collect a sample at the same
place. b.
The following are the most common locations for collecting a sample at a
facility. They may or may not be described in the permit.
1.
Effluent - Effluent samples should be collected at the site specified in
the permit, or if no site is specified in the permit, at the most representative
site downstream from all entering wastewater streams prior to discharge to the
appropriate disposal method (e.g., surface water discharge, ground water
discharge, wetlands discharge, deep-well injection).
2.
Influent - Influent wastewaters are preferably sampled at points of
highly turbulent flow in order to ensure good mixing;
however, in many instances the most desirable location may not be
accessible. Preferable influent
wastewater sampling points include: a.
the upflow siphon following a comminutor (in absence of grit chamber); b.
the upflow distribution box following pumping from main plant wet well; c.
aerated grit chamber; d.
flume throat; e.
pump wet well when the pump is operating.
In all cases, influent samples shall be
collected upstream from recirculated plant supernatant and residuals, and the
sample collected should be completely untreated.
3.
Internal Outfalls - Internal outfalls are to be sampled as specified in
the permit or consent order. a.
This type of sampling occurs on special projects, on industries or
domestic facilities undergoing a diagnostic inspection, on treatment trains in a
facility which are essential to the final effluent quality, and on facilities
which occupy watersheds (e.g., phosphate mines). b.
Sampling of internal outfalls is uncommon, but can be useful in
identifying potential equipment problems and unidentified contaminant sources.
4.
Groundwater Monitoring Wells - Ground water monitoring wells should be
sampled as described in the Groundwater Section.
5.
Groundwater Discharge Sites - Sample collection of discharges to land
application sites and underground injection wells are best taken at the effluent
sampling points.
6.
Residuals - Residual samples should be taken in accordance with EPA's
POTW Sludge Sampling and Analysis Guidance Document, 1989.
7.
Residual Sites - Samples taken at a residuals disposal site are taken to
determine build-up of pollutants in the soils.
8.
Pond and Lagoon Sampling - Generally, composite wastewater samples should
be collected from ponds and lagoons. Even
if the ponds or lagoons have long retention times, composite sampling is
necessary because of the tendency of ponds and lagoons to stratify.
However, if dye studies or facility data indicate a homogeneous
discharge, a grab sample may be taken as representative of the waste stream.
9.
Surface Water Sites - The location of surface water sampling sites for
wastewater facilities depend on whether the facility discharges to a
unidirectional flowing body of water (e.g., stream, river) or a
non-unidirectionally flowing body of water (e.g., tidally influenced coastal
rivers, bays, estuaries, lakes). a.
Unidirectional Flow Streams - The upstream control site, if one exists,
should be just far enough upstream to be out of the influence of any effluent.
This may entail going farther upstream to avoid any potential groundwater
contamination from adjacent spray irrigation sites or percolation ponds.
1. The
downstream or test site should be at the edge of any mixing zone (if there is
one) and far enough downstream to be in the peak zone of impact.
These sites should be determined on a case-by-case basis.
2.
The control and test sites should be matched as closely as possible on
habitat structure (e.g., flora, pool/riffle type, shading) based on a habitat
analysis.
3.
If there is no upstream site (e.g., when the discharger currently forms
the upstream flow), then a reference site should be chosen in a nearby stream
based on a habitat analysis. b.
Nonunidirectional Flow Water Bodies - For tidally influenced rivers,
bays, estuaries, swamps, lakes, ponds, and other water bodies which don't have
unidirectional flow, an unimpacted reference site must be selected.
1.
The site should be matched carefully on habitat structure (e.g., type,
sediments, stream order/size, type of drainage) to ensure you are comparing the
same type of sites.
2.
The outfall should have two test sites either at the edge of the mixing
zone or in the predicted zone of impact. The
two sites should be located in different directions depending on where the
effluent plume would be expected during tidal changes, in the direction of
prevailing wind, or along anticipated flow gradients even though undetectable.
10.
Effluent limits in a permit are often specified as a mass loading.
To determine a mass loading and thereby evaluate compliance with permit
limits, it is necessary for the sampler to obtain accurate flow data.
Flow measurement is the commonly used term for this process.
In addition to verifying compliance with permit limits, flow measurement
serves to: a.
Provide operating and performance data on the wastewater treatment plant b.
Compute treatment costs, based on wastewater volume c.
Obtain data for long-term planning of plant capacity.
Specific operating instructions for automatic samplers, capabilities,
capacities, and other pertinent information are included in the respective
operating manuals and are not presented here.]] 4.2.4.3
Sample Types There
are two primary types of samples: 1)
grab samples; and 2) composite samples. Each
type has distinct advantages and disadvantages.
In order to obtain a more complete characterization of a specific
facility's effluent, the two sample types can be used independently or in
combination. a.
Grab Samples
1.
This is an individual sample collected over a period of time, usually all
in one motion, generally not exceeding 15 minutes.
The 15 minute time limit applies to aqueous samples only.
No particular time limit applies to the
collection of solid samples (e.g. residuals). Grab samples may be used to determine consistency between an
industry's self-monitoring data and to corroborate the results of composite
samples.
2.
Grab samples represent the conditions that exist at the moment the sample
is collected and do not necessarily represent conditions at any other time.
Grab sampling is the preferred method of sampling under the following
conditions: a.
A snapshot of the wastewater quality at a particular instant in time is
desired; and b.
The water or wastewater stream is not continuous (e.g., batch discharges
or intermittent flow); c.
The characteristics of the water or waste stream are known to be constant
or nearly so; d.
When the waste conditions are relatively constant over the period of
discharge. In lieu of complex sampling activities, a grab sample
provides a simple and accurate method of establishing waste characteristics; e.
The sample is to be analyzed for parameters whose characteristics are
likely to change significantly with time (i.e., dissolved gases, bacteria, pH,
etc.); f.
The sample is to be collected for analysis of a parameter such as oil and
grease or bacteriologicals where the compositing process could significantly
affect the actual concentration; g.
Data on maximum/minimum concentrations are desired for a continuous water
or wastewater stream; and h.
Identifying and tracking slug loads and spills.
3.
If required to be measured, the following parameters shall be measured on
grab samples or in-situ. NOTE:
If the permit specifies a composite sample for any of the above-mentioned
parameters, THE PERMIT CONDITIONS SHALL BE FOLLOWED.
4.
Sampling protocols shall follow those outlined under Surface Water
(4.2.3.3.a). b.
Composite Samples
1.
A composite sample is a sample collected over time, formed either by
continuous sampling or by mixing discrete samples.
Composite samples reflect the average characteristics during the
compositing period.
2.
Composite samples are used when stipulated in a permit and when: a.
The water or wastewater stream is continuous; b.
Analytical capabilities are limited; c.
Determining average pollutant concentration during the compositing
period; d.
Calculating mass/unit time loadings; and e.
Associating average flow data to parameter concentrations.
3.
Composite samples may be collected individually at equal time intervals
if the flow rate of the sample stream does not vary more than plus or minus ten
percent of the average flow rate, or they may be collected proportional to the
flow rate. The permit may specify which composite sample to use, either
time composites or flow proportional composites.
The compositing methods, all of which depend on either continuous or
periodic sampling, are described in the following discussions.
4.
Time Composite Sample a.
Time composite samples are based on a constant time interval between
samples. b.
A time composite sample can be collected manually or with an automatic
sampler. c.
This type of composite is composed of discrete sample aliquots collected
in one container at constant time intervals. d.
This method provides representative samples when the flow of the sampled
wastewater stream is constant. This
type of sample is similar to a sequential composite sample (described below).
5.
Flow Proportional Composite Sample a.
Flow proportional samples can be collected automatically with an
automatic sampler and a compatible pacing flow measuring device,
semi-automatically with a flow chart and an automatic sampler capable of
collecting discrete samples, or manually. b.
There are two methods used to collect this type of sample:
1.
One method collects a constant sample volume per stream flow (e.g., 200
milliliters (ml) sample collected for every 5,000 gallons of stream flow) at
time intervals proportional to stream flow.
This method provides representative samples of all waste streams when the
flow is measured accurately. For
this reason, it is used frequently.
2.
In the other method, the sample is collected by increasing the volume of
each aliquot as the flow increases, while maintaining a constant time interval
between the aliquots (e.g., hourly samples are taken with the sample volume
being proportional to the flow at the time the sample is taken).
6.
Sequential Composite Sample - Composed of discrete samples taken into
individual containers at constant time intervals or constant discharge
increments. a.
For example, samples collected every 15 minutes are composited for each
hour. b.
The 24-hour composite is made up from the individual one-hour composites.
1.
Each of the 24 individual samples is manually flow proportioned according
to the flow recorded for the hour that the sample represents.
2.
Each flow proportioned sample is then added to the composite samples.
3.
The actual compositing of the samples is done by hand and may be done in
the field or the laboratory.
4.
In most cases, compositing in the field is preferable since only one
sample container must be cooled, and then transported to and then handled in the
laboratory.
5.
A 24-hour composite is frequently used since an automatic sampler can
easily collect the individual samples. c.
A variation of the 24-hour composite is to collect a constant volume of
sample taken at constant discharge increments, which are measured with a
totalizer. For example, one aliquot is collected for every 10,000
gallons of flow. d.
Sequential sampling is useful to characterize the waste stream because
you can determine the variability of the wastewater constituents over a daily
period. For example, for
pretreatment studies you can visually determine when high strength wastes are
being discharged to a facility or when heavy solid
loads are being discharged during a 24-hour cycle.
You can measure different pHs throughout the day.
The value of this type of sampling must be weighed against the manpower
constraints and sampling goals.
7.
Continuous Composite Sample - Collected continuously from the waste
stream. The sample may be a
constant volume which is similar to the time composite, or the volume may vary
in proportion to the flow rate of the waste stream, in which case the sample is
similar to the flow proportional composite.
8.
Areal Composite - A sample composited from individual grab samples
collected on an areal or cross-sectional basis.
Areal composites shall be made up of equal volumes of grab samples; each
grab sample shall be collected in an identical manner.
Examples include soil or residual samples from grid system points on a
land application site, water samples collected at various depths at the same
point or from quarter points in a stream, etc. 4.2.4.4
General Concerns a.
The sampler must weigh advantages and disadvantages when choosing between
the use of grab or composite sampling methods.
1.
While grab sampling allows observation of unusual conditions that may
exist during discharge, such as sudden bursts of color or turbidity, this method
is labor intensive and impractical when sampling is performed at many locations
over extended periods of time.
2.
When sampling a large number of locations, the use of automatic samplers
is more practical. a.
Automatic samplers also help reduce human error, specifically in complex
sampling activities, such as flow proportional sampling, and reduce exposure to
potentially hazardous environments. b.
The primary disadvantage to automatic sampling is the cost of the
equipment and maintenance requirements. Many
automatic samplers in use today are electronically controlled and must be sent
back to the manufacturer when a malfunction occurs. b.
In order to obtain a representative sample, sampling must be conducted
where wastewater flow is adequately mixed.
In general these criteria shall be used to evaluate the location:
1.
A sample should be taken in the center of the flow where velocity is
highest and there is little possibility of solids settling. a.
The sample should be collected at a depth between 40% - 60% of the total
depth where the turbulence is maximized. This
means that sample collection should be avoided at the water surface or the
channel bottom. b.
Flow mixing is particularly important for ensuring uniformity.
2. Sampling
personnel should be cautious when collecting samples near a weir because solids
tend to collect upstream and floating oil and grease accumulate downstream.
3.
If the sample is not to be tested for volatile organics or will not be
affected by stripping of dissolved gases, mechanical stirring may be used or a
stream of air may be introduced into the waste stream.
4.
In sampling from wide conduits, cross-sectional sampling should be
considered. Dye may be used as an
aid in determining the most representative sampling point(s).
Note: appropriate Department personnel should be consulted for the type
of dye and acceptable protocols.
5.
If manual compositing is employed, the individual sample bottles must be
thoroughly mixed before pouring the individual aliquots into the final composite
container.
6.
f the sample is taken from an effluent tap, allow the tap to run for one
- two minutes to allow the settled solids to flush from the line.
Reduce the flow to 500 ml/min before collecting the samples. c.
Sampling and flow measuring are integrally related.
The sampler must know the wastewater flow variability before a sampling
program can be initiated. Whether
to use a flow proportional or time composite sampling scheme depends on the
variability of the wastewater flow. If
a sampler knows or suspects significant variability in the wastewater flow or
knows nothing about the facility, a flow proportional sample should be
collected; otherwise a time composite sample would be acceptable. d.
Prior to sampling, the flow measuring system (primary flow device,
totalizer, recorder) should be examined (see sections on Stage and Flow
Measurement in the EPA SOP, Sections 7.6 and 7.8).
If the flow measuring system is unacceptable, the sampler may have to
install a flow measuring device. If
the flow measuring system is acceptable, samples can be collected by the
appropriate method. e.
Fill out the information on the sample container tags and on the field
sheets completely and carefully. Improper
sample identification results in invalid or unacceptable samples and lost
sampling efforts. f.
Take inordinate care to prevent cross-contaminating samples.
Use properly cleaned sampling equipment. 4.2.4.5
Sample Equipment Requirements a.
Manual Sampling
1.
The types of sampling devices that may be used to collect samples are
specified in the next section. Additional
discussions are found the protocols for collection grab samples in surface water
(section 4.2.3.3.a).
2.
IN ALL CASES, the selected sampling equipment shall be compatible with
the components to be collected and shall comply with the use and construction
restrictions specified in Table 4.1.
3.
All equipment shall be cleaned using the appropriate protocols specified
in Section 4.1. Sample containers shall be cleaned according to Section 4.4.1
or obtained precleaned from commercial sources.
b.
Automatic Samplers
1.
A wide variety of automatic samplers are commercially available (e.g.,
Sigma, ISCO). Most have the
following five interrelated subsystem components: a.
Sample Intake Subsystem - The sample intake gathers representative
samples from the sampling stream.
1. The
intake is usually the end of a plastic suction tube which should also be
resistant to physical damage from large objects in the flow stream.
Nonleaching tygon tubing is most often used.
2.
Teflon tubing shall be used under the conditions specified in
4.2.4.5.b.2.b. The end of this
tubing should be fixed to a piece of conduit or a pole bent to hold the sample
port in the waste stream at the correct location to get a representative sample.
The tubing shall be supported in such a way that the incoming sample is
not contaminated by either the supporting pole or the method of attachment. b.
Sample Gathering Subsystem - Automatic samplers provide one of three
basic gathering methods:
1.
Mechanical - Mechanical gathering subsystems are usually built into place
and include devices such wide/deep channel flow.
Because of the mechanical as cups on cables, calibrated scoops, and
paddle wheels with cups. Although
these systems obstruct the stream flow, they take into account site specific
considerations, such as high sampling lift and system employed, these units
require periodic maintenance.
2.
Forced Flow - Forced flow gathering subsystems are often built into place
as permanent sampling facilities; thus, like the mechanical gathering
subsystems, they may obstruct the stream flow.
They also require periodic inspection and maintenance.
However, forced flow subsystems have the advantage of being able to
sample at great depths. In
addition, because this gathering system uses air pressure to transport the
sample, it may be ideal for sample collection in potentially explosive
environments.
3.
Suction Lift - The suction lift is the most widely used type of sample
gathering subsystem due to its versatility and minimal affect on flow patterns.
Suction lifts are limited to 25 vertical feet or less because of internal
friction losses and atmospheric pressure. At
least 100 ml should be collected each time the pump is actuated. c.
Sample Transport Subsystem - The sample is usually transported from the
sample intake to the collection bottle by a plastic tube referred to as the
sample transport subsystem. Care
should be exercised to avoid sharp bends and twists in the transport line. d.
Sample Storage Subsystem - The sample storage subsystem can accommodate
either a single large collection bottle or a number of smaller collection
bottles.
1.
The total sample volume storage capability should be at least 2 gallons
(7.6 liters): some samplers have a
capacity as great as 5 gallons.
2.
To thermally preserve the samples, storage subsystems must be large
enough to provide space for ice to chill the sample during collection (see b.3
below).
3.
Samples with individual bottles for discrete collection are usually
equipped with a cassette which rotates to fill the bottle at the time of
sampling. e.
Controls and Power Subsystem
1.
The control units allow selection of time or flow compositing method, or
continuous sampling method. The
automatic samplers most widely used have encapsulated solid state controls.
This minimizes the effects of unfavorable environments that may be
encountered in the field, such as high humidity and corrosiveness. These units are also sealed so that they may be used with
minimum risks in potentially explosive environments.
In addition, sealed units protect the controls if the sampler is
accidentally submerged.
2.
Samplers operating from a power supply are more reliable than battery
operated models; however, field conditions often prohibit the use of a power
supply.
2.
Automatic sampling equipment must meet the following requirements: a.
Sampling equipment must be properly cleaned to avoid cross contamination
which could result from prior use (see section 4.1.5 for specific cleaning
procedures). b.
If samples for organics (includes all extractable organics, pesticides,
and herbicides and TOC) are to be collected, no plastic or non stainless steel
parts of the sampler shall come in contact with the water or wastewater stream:
1.
Teflon tubing shall be used in the transport subsystem.
2.
A special sampler base and glass containers may also be necessary for
sampling organics. Consult your
owner's manual. c.
If the preservation requirements for a particular component specify that
a sample be thermally preserved, the sampler must be able to keep the samples
cool to 4 C during the sampling period. This
is accomplished in the field by using ice or refrigeration units in the sampler. d.
The sampler must be able to collect a large enough sample for all
parameter analyses. Additionally, split samples may also be necessary. e.
A minimum of 100 ml should be collected each time the sampler is
activated, if a peristaltic pump is used. f.
The sampler should provide a lift up to at least 20 feet and the sampler
should be adjustable so that volume is not a function of the pumping head. g.
The pumping velocity must be adequate to transport solids and not allow
solids to settle. h.
Pump intake line
1.
The automatic sampler must provide for line purging after each sample is
drawn to prevent contamination of subsequent samples.
2.
The minimum intake line inside diameter shall be at least 1/4 inch, which
is large enough to lessen chances of clogging but small enough to maintain
velocity and to avoid solids settling. i.
Sample transport system
1.
The tubing should be at least 1/4 inch inside diameter to maintain
adequate flow and to prevent plugging.
2.
Tubing should be sized so that a velocity of at least two feet per second
can be maintained.
3.
Line must be automatically purged after each sample is collected. j.
An adequate power source should be available to operate the sampler for
48 hours at a 30-minute sampling interval. k.
Sample collection vessels large composite or discrete sample containers,
shall be constructed of materials appropriate for the tests to be performed.
As a general rule, the vessels shall be made of the same material as
those specified for sample containers in section 4.4.2.3.
3. In
addition to the requirements listed in 4.2.4.5.b.2 above, several factors should
be considered in selecting automatic sampling equipment.
Among these are: a.
Convenience of installation and maintenance - Sampling equipment should
always be handled carefully and maintained in accordance with the manufacturer's
instructions. Most equipment
failures are caused by careless handling and poor maintenance. b.
Equipment security
1.
Security is important when sampling is done as part of an enforcement
proceeding.
2.
Manhole locations where battery operated equipment may be installed and
the cover replaced will aid in maintaining security.
3.
If sampling equipment must be left unattended, the sampler should be
provided with a lock or seal which, if broken or disturbed, would indicate that
tampering had occurred. c.
Operation in cold or hot weather
1.
Cold weather - In Florida, use of automatic samplers is seldom a problem
during cold weather. If a sample
must be taken during extremely cold days freezing of intake lines may happen.
These problems may be handled by using heat tape or placing the sampler
inside a thermostatically controlled, electrically heated enclosure.
In the absence of special equipment, freezing may be prevented by
installing the sampler in a manhole or wet well or by wrapping the sampler with
eight or nine inches of insulation and wind protection.
Also, the sampler should be positioned well above the effluent stream so
that the tubing runs in a taut, straight line to prevent pooling of liquid.
2.
Hot weather - The summer heat in Florida does pose a significant problem
with keeping the sample cool. If
possible, choose a shaded or even cooled place for the sampler.
If not, insulation wrapped around the sampler may help. Painting the sampler white will reflect some heat.
Before leaving the site refill the automatic sampler with ice or check to
see that the refrigeration unit is operating.
An attempt should be made to pick up the samples near the time the last
sample is taken. Samples which are allowed to warm up are questionable, if not
useless. 4.2.4.6
Sample Acquisition a.
Manual Sampling
1.
Manual sampling is used for collecting grab samples for immediate in-situ
field analyses. However, it can also be used in lieu of automatic equipment
over extended periods of time for composite sampling, especially when it is
necessary to observe and/or note unusual waste stream conditions.
2. Collection
using the Sample Container a.
The actual sample container must always be used for collecting samples
for oil and grease, volatile organic compound (VOC), and bacteriological
samples. b.
If possible, manually collected samples should be collected in the actual
sample container that will be submitted to the laboratory.
This eliminates the possibility of contaminating the sample with an
intermediate collection container. c.
Sample containers containing premeasured amount of preservatives SHALL
NOT be used to collect surface water grab samples.
Alternative collection procedures listed below must be followed. d.
Sample collection shall follow the protocols for collecting simple grab
samples in surface water (Section 4.2.3.3.a).
3.
Sampling with an intermediate vessel or container a.
If the sample cannot be collected as described above, an intermediate
vessel can be used. b.
The sample shall be collected following the protocols described for
collecting surface water grab samples with an intermediate container or pond
sampler (Section 4.2.3.3.a) and redistributed into appropriate sample
container(s).
4.
Samples collected in bailers a.
Bailers may be used if the data requirements do not necessitate a sample
from a strictly discrete interval of the water column. b.
Bailers with an upper and lower check-valve can be lowered through the
water column and water will be continually displaced through the bailer until
the desired depth is reached, at which point the bailer is retrieved.
This technique may not be successful in strong currents. c.
Specific sampling protocols outlined under mid depth surface water
sampling (4.2.3.3.b.4) shall be followed. 5.
Samples collected with pumps a.
In some cases it may be best to use a pump, either power or hand
operated, to withdraw a sample from the water or wastewater stream. b.
Protocols for the use of pumps are specified in the surface water
sampling Section (4.2.3.3.b.5) and shall be followed when collecting with pumps.
6. Dedicated
equipment may also be used at each sampling station.
This will avoid cross contamination between sampling stations.
For most parameters (other than trace metals and organics) rinsing the
sampling device three times in the effluent stream is sufficient.
More stringent requirements must be used for trace pollutants (see
Section 4.1). b.
Automatic Samplers
1.
Automatic samplers may be used when several sites are to be sampled at
frequent intervals or when a continuous sample is required.
2. Conventional
Sampling: All composite samplers
can be used to collect time composite or flow proportional samples. a.
In the flow proportional mode, some samplers are activated by a
compatible flow meter. b.
Flow meter operation will not be discussed here. Refer to the operating
manuals if you have them. c.
For older models, flow proportional samples can be collected using a
discrete sampler and a flow recorder and manually compositing the individual
aliquots in flow proportional amounts.
3.
Installing the Composite Sampler a.
All new or precleaned tubing (Dow Corning Medical Grade Silastic, or
equal, in the pump and either Teflon or Tygon, depending on the parameters of
interest, in the sample train) shall be used each time the sampler is installed.
1.
Cut the proper length of precleaned Teflon or Tygon tubing;
2.
Rinse deionized water through the sampler and collect an equipment blank;
3.
Put the collection sieve and tubing in the appropriate sample location in
the wastewater stream, using conduit if necessary to hold it in place.
Assure that the incoming sample water is not contaminated by the
supporting conduit; and
4.
Program the sampler as per manufacturer's directions and as required in
the permit conditions. b.
For a time composite sample, the sampler should be programmed to collect
200 ml at 30-minute intervals or 100 ml at 15-minute intervals into a
refrigerated 3-gallon jug. For a
5-gallon compositing jug, the volume should be increased accordingly. c.
For a flow proportional sample, the sampler should be programmed to
collect a minimum of 100 ml for each sample interval, with the interval
predetermined based on the flow of the waste stream. d.
At the end of each 24-hour sampling period, the contents of the
compositing jug (sample) should be stirred and siphoned (poured if no visible
solids) into the respective containers, followed by immediate preservation, if
required.
4. Automatic
Sampler Security - A lock or seal
may be placed on the sampler to prevent or detect tampering.
However, this does not prevent tampering with the sampler tubing (see
additional discussions on sample security under equipment requirements
4.2.4.5.b.3 above)
5. Sample
Preservation - Samples shall be preserved for all samples according to 40 CFR
Part 136 Table II. a.
Table II includes allowances for automatic samplers.
In addition to the capability of keeping samples cooled with ice or
refrigeration, there are 2 considerations to be presented for chemical
preservation:
1.
If separate bottles are used, they may be prepreserved with the
appropriate chemical preservative or preserved after sampling has been completed
(WITHIN 24 HOURS);
2.
If the large compositing jug is used, preservation should be completed
after sampling has been completed (WITHIN 24 HOURS);
3.
NOTE: If the only parameter of interest is Total Phosphorus, and the
project is unrelated to a NPDES permit, then the sample must be chemically
preserved (H2SO4) but it need not be cooled to 4 C with wet ice. The
acid must be in the container prior to sample collection. 4.2.5.1
Introduction and Scope This section presents the standard
operating procedures that should be employed during field investigations to
ensure that representative groundwater samples are collected.
The importance of proper sampling of monitor wells cannot be
overemphasized. Care should be
taken so that the sample collected is neither altered nor contaminated by
sampling and handling procedures. The following discussions cover
acceptable: equipment choice, equipment construction materials, pre-sampling and
in-field decontamination, purging and sampling technique, and proper field
Quality Control procedures. Although
not a complete discussion of all groundwater sampling procedures, this
information has been compiled
with the intent of providing the equipment and techniques for situations that
are most likely to be encountered. [[4.2.5.2
Selection of Sampling Materials This section is intended as a GUIDE to
be used when selecting sample collection and/or well purging equipment. The discussions are informational and should be considered as
criteria when selecting equipment. The selection of inert materials used
for well purging, sample collection, handling and storage is a critical
consideration in planning the well-conceived cost-effective monitoring program. These equipment descriptions may also be used for other
sampling matrices. The materials of
choice should retain their structural integrity for the duration of the
monitoring program. They must not
react with the sample (i.e., should
not absorb, adsorb or leach sample constituents) which would bias representative
sample collection. Additionally,
the sampling equipment should be chosen so that they can be easily
decontaminated and transported. The
following subsections, adapted from Barcelona, et al. (1983) and Watts (1988),
discuss sampling material construction. a.
Stainless Steel - The most common types of stainless steel that are used
in sample collection equipment are 316 and 304 although other types are
available and may be used.
1.
Stainless Steel-316 a.
Recommended for use in most groundwater and soil monitoring situations
for all parameter groups. b.
Stainless Steel-316 should be utilized especially for detailed organic
and trace metals analytical work. c.
Stainless Steel-316 is least likely to introduce sampling bias or
imprecision. d.
Stainless Steel-316 is also relatively easy to decontaminate without
requiring a HNO3 rinse. e.
The cost per foot of Stainless Steel-316 is approximately the same as
Teflon. f.
Note that dedicated stainless steel sampling equipment (tubing,
submersible pump housing, etc.) may not be suitable to collect samples for
metals analyses since its performance may be sensitive to the chloride ions,
which can cause pitting and corrosion over long term exposure under acidic
conditions.
2.
Stainless Steel-304 a.
Recommended for use in most groundwater and soil monitoring situations
for all parameter groups (including trace organics and metals).
b.
Stainless Steel-304 is less corrosion-resistant than 316 and may be prone
to show pitting and corrosion if left in contact with acidic high total
dissolved solids (TDS) groundwaters for extended periods (dedicated downhole
equipment). c.
Corrosion products are mainly limited to iron and potentially Cr, Cu and
Ni. Pitted surfaces may present
active adsorption sites and render difficulty in future decontamination. c.
Teflon
1.
Teflon is the trademark of Dupont, Inc.
Teflon is recommended for use in most groundwater and soil monitoring
situations for all parameter groups, especially for detailed, trace organic
analytical work.
2.
Teflon material is least likely to introduce sampling bias or
imprecision. Teflon is relatively
easy to decontaminate. Teflon is a relatively soft material and is easily scratched.
3.
Bailers and tubing that are old and excessively scratched will be
difficult to clean and must be discarded.
4.
Some Teflon bailers have replaceable "donuts" that encircle the
bailer and protect the body from scratches. d.
Low-Carbon Steel, Galvanized Steel and Carbon Steel
1.
These materials are generally not recommended for collecting samples,
especially galvanized steel.
2.
In cases of split spoon and core barrel soil sampling these might be the
only material available. a.
If trace metals are of interest, plastic or teflon liners shall be used
to prevent the sample from contacting the equipment surface. b.
Teflon liners are recommended if organics are of interest.
3.
New equipment made from these materials must be very carefully cleaned to
remove oily manufacturing residues.
4.
Corrosion is likely in high dissolved solids and acidic environments,
particularly when sulfides are present.
5.
Products of corrosion for carbon steel are mainly Fe and Mn, except for
galvanized steel which may release Zn and Cd.
6.
Weathered steel surfaces present very active adsorption sites for trace
organic and inorganic chemical species and pitted surfaces will increase the
difficulty of decontamination. e.
Polypropylene/Polyethylene
1.
Polypropylene and polyethylene are polyolefin materials that are more
resistant to organic solvent attack than formulated plastics such as Viton,
silicone or neoprene.
2.
Polypropylene and polyethylene are comparable in performance and
resistance to Teflon in corrosive high dissolved solids (Pettyjohn et al., 1981
and Barcelona et al., 1983).
3.
Polypropylene and high density polyethylene (HDPE) tubing material may be
used for purging wells (see Table 4.1 for details).
4.
HDPE disposable bailers may be used to sample all parameters except
organic compounds. f.
Polyvinyl chloride (PVC)
PVC (flexible) is not recommended for detailed
organic analytical work. The
polymer formulated PVC contains plasticizers, stabilizers and antioxidants which
may cause interferences with analytical determinations, especially when
industrial solvents are encountered in groundwater.
Documented interferences are likely with several priority pollutant
classes. g.
Viton, Tygon, Silicone and Neoprene
Viton, Tygon, silicone and neoprene are not
recommended for organic analytical work since the inherent plasticizers,
stabilizers and antioxidants may introduce interferences.]] 4.2.5.3
Purging and Sampling Equipment a.
General Considerations
1.
Purging the monitor well of stagnant water can be performed with various
equipment. The choice of equipment
will depend on the parameters of interest, the well diameter, the well specific
capacity, transmissivity, the water level elevation and other site conditions.
As stated earlier, the choice of equipment used for purging must not bias
the "representativeness" of the sample collected.
2.
It is recommended that field personnel use pumps to purge monitor wells
if at all possible.
3.
Bailers are not recommended for purging monitor wells because frequent
lowering and retrieving of the bailer: a.
will introduce atmospheric oxygen which may precipitate metals (e.g.
iron) or cause other changes in the chemistry of the formation water (i.e. pH), b.
will result in agitation or volatilization of groundwater which may bias
volatile and semi-volatile analyses, and c.
may introduce dirt through scraping the sides of the casing wall.
4.
Though bailers are not recommended for purging, they are acceptable if
constructed of the appropriate material and if extreme care is used.
The use of bailers is described in 4.2.5.3.c below.
5.
All standing water around the top of the well casing (manhole) shall be
removed before opening the well. b.
Pumps - A summary of the principles of operation and
the advantages and disadvantages of the various commercial pumps is given
in "Monitor Well Construction and Groundwater Sampling" (Watts, 1988).
1.
Above-ground Pumps a.
Peristaltic Pump - Peristaltic pumps may be used to purge low volume, low
specific capacity wells in which the static water level in the well is no
greater than 20-25 feet BLS (Below Land Surface).
1.
Decreased pumping velocity will be experienced when water levels are
deeper than 18'-20'.
2.
It also may be used to sample wells for limited parameter groups.
These parameter groups will be dependent upon tubing materials and
arrangements. It is the preferred
method of collecting filtered groundwater samples for metals.
See Table 4.1 for details on the restrictions for this pump, including
choice of tubing (i.e. Teflon, HDPE, Tygon).
See EPA Region IV SOP & QAP, Appendix F.1 for additional guidance. b.
Centrifugal Pump - Centrifugal pumps can be utilized to purge 2 inch and
larger internal diameter wells that have moderate specific capacities from 2 -
10 gpm (gallons per minute) and have a static water level greater than 20 feet
BLS.
1.
The pump may also be attached directly to 3/4" well point casing and
used to purge (see 4.2.8, care must be taken so that purged water does not fall
back into the well casing).
2.
Sampling gloves shall be worn and discarded after positioning the pump.
Hands should be washed and new gloves shall be put on prior to sampling.
3.
See Table 4.1 for compatibility restrictions related to choice of tubing
and allowable parameter groups.
2.
Submersible Pumps a.
Electric Submersible Pumps - Submersible pumps (e.g. Grundfos, Goulds,
Jacuzzi) can be utilized for purging 4 inch or greater diameter monitor wells.
Some submersible pumps can be utilized in 2 inch wells (e.g. Fultz and
Grundfos). These pumps can be used
in wells that have moderate to high specific capacity and cannot be purged using
an above-ground pump because of the lower static water level elevation
(>20'-25' BLS).
1.
The pump must be constructed of stainless steel (and/or Teflon) material
and the delivery hose shall be constructed of appropriate material depending
upon the analytes of interest.
2.
It may be fitted with inert stainless steel or Teflon tubing between the
pump and "other non-inert tubing" to be able to purge wells that will
be sampled for trace organics.
3.
See EPA Region IV SOP & QAM, Appendices F.2-F.3 for further
information regarding 2 and 4 inch electric, submersible pumps. b.
Bladder Pumps - Positive-displacement bladder pumps (no-gas contact) can
be utilized for purging wells where the water table is greater than 25 feet and
an above-ground pump cannot be used. These
pumps are used in wells with low to moderate capacity since pumping rates are
not as high as the electric submersibles or the gas-contact "purge
pump" described below. Maximum
pumping rates are approximately 0.5 - 1.5 gallons per minute depending upon the
location of the pump (BLS).
1.
The bladder pump system is composed of three major components:
the pump, the compressed air and water discharge tubing, and the
controller/compressor.
2.
The pump consists of a bladder and an exterior casing or pump body that
surrounds the bladder. These two
parts can be composed of various materials, usually combinations of PVC, Teflon,
and stainless steel.
3.
The construction material of the pump body, pump bladder, and the
discharge tubing will define the parameters that can be purged and sampled with
this system.
4.
If the pump is not permanently installed in the monitor well and if it is
to be used to purge and/or sample for all parameters (including VOCs), the pump,
bladder and tubing must be constructed of stainless steel and Teflon.
5.
Permanently installed pumps have a PVC pump body as long as the pump
remains in contact with the formation water.
If VOCs and/or extractable organics are of interest, the bladder and the
delivery tubing shall be constructed of teflon. c.
Bladderless Purge Pumps - These pumps are identical to the bladder pumps
described above except they do not have an internal bladder.
The air controller/compressor is used to force water from within the pump
body up the discharge tubing. By
not having the (Teflon) bladder fill by head pressure, pumping rates are much
higher (>4 gpm).
1.
This pump can only be used for purging.
2.
Additionally, operation of this pump cannot result in purge water
escaping back into the well. Proper
operation and maintenance of the check valve must be ensured.
Release of aerated purge water into the water column is not acceptable.
3.
Hand Pumps a.
Hand pumps (e.g. Brainard-Kilman 'B-K Pump') are manual pumps that should
be utilized for purging 2- or 4-inch diameter monitor wells in which the static
water level is too deep for use of a centrifugal or peristaltic pump.
1.
The B-K hand pump and the associated riser pipes are constructed of PVC
and shall be used to purge when only inorganic constituents are of interest
unless the restrictions specified in Table 4.1 are followed.
4.
The lower most section of the B-K pump is equipped with a foot valve to
prevent back flow of purge water.
5.
After purging has been completed, the B-K pump should be completely
disassembled and decontaminated.
6.
Please see Table 4.1 for details on the use of this pump. c.
Bailers
1.
As stated above, the use of bailers is not recommended for purging.
2.
Bailers shall be composed of material compatible with the analytes of
interest. See Table 4-1 for
restrictions a.
Bailers constructed of stainless steel and Teflon may be used to sample
any and all parameters. b.
Bailers constructed of high density (rigid) polyethylene (HDPE) materials
may be used to sample monitor wells for inorganics and free-product only. c.
When sampling grossly contaminated tanks or other facilities, disposable
polyethylene (or other material) bailers should be utilized (it may be difficult
to decontaminate such grossly contaminated bailers and as such they may have to
be discarded).
3.
The bailer must be handled carefully so as not to contaminate it prior to
use.
4.
They shall be scrupulously cleaned, including all check valves (see
Section 4.1.4.1). d.
Lanyards
1.
Lanyard may be disposable (braided or monofilament nylon or reusable
(stainless steel or teflon-coated).
2.
A disposable lanyard must be changed for each monitor well, but the same
lanyard may be used for purging (if performed) and sampling operations without
decontamination between purging and sampling operations.
3.
Reusable lanyards shall be decontaminated between monitor wells (see
Section 4.1.9.1) but do not require cleaning between purging and sampling
operations. 4.2.5.4
Water Level and Purge Volume Determination Prior to sampling, an adequate amount
of stagnant well water in the well must be removed in order to sample
representative formation water. a.
Inspect the exterior protective casing monitor well for damage and
document accordingly. b.
Water Level Measurements
1.
In order to calculate the purge volume, the water level is determined by
using an electronic probe, chalked tape, etc.
2.
The depth below land surface shall always be recorded to the nearest 0.1
foot from the same reference or survey mark on the well casing.
3.
Measurements using an electronic probe shall follow the manufacturer's
instructions. Since false reading
may be obtained if the sensor contacts the well casing, multiple readings shall
be taken to ensure accuracy.
4.
Chalked Tape Method a.
Lower chalked tape into the well until the lower end is in the water
(usually determined by the sound of the weight hitting the water). b.
Record the length of the tape relative to the reference point (see 2
above). c.
Remove tape and note the length of the wetted portion. d.
The depth to water is determined by subtracting the length of the wetted
portion (c above) from the total length (b. above).
5.
Decontaminate all measuring devices immediately after use (see 4.1.9.1)
and prior to next measurement. c.
Water Column Determination
1.
The total water column is obtained by subtracting the depth to the top of
the water column from the total depth of the well.
2.
Total depth of well is dependent upon the well construction.
Some wells may be drilled in areas of sinkhole or karst formations. In cases where there may be an open borehole below the cased
portion, an attempt should be made to find the total borehole depth. d.
Well Water Volume
The length of the water column is then
converted to volume of water that is present in the well:
1.
2 inch casing:
V = 0.17 X h
Where: V
= volume in gallons
h
= height of the water column in feet
2.
4 inch casing:
V = 0.66 X h
Where: V
= volume in gallons
h
= height of the water column in feet
3.
For other casing sizes, calculate using the following:
V = (0.041)d X d X h
Where: V
= volume in gallons
d
= well diameter in inches
h
= the height of the water column in feet
or:
V = Pr X r X h(0.001)
Where: V
= volume in liters
P
= 3.14159 (pi)
r
= radius in centimeters
h
= height of water column in centimeters e.
Record all measurements in the field records. 4.2.5.5
Well Purging Techniques To ensure a representative groundwater
sample from a monitor well it is essential that the well be purged prior to
sampling. Stagnant water in a well
casing may undergo a variety chemical changes due to alterations in the redox
potential, pH and leaching of organic compounds from the casing. Several
methods of purging wells have been cited in the literature (Gibb et al., 1981,
and Barcelona et al., 1983). The
choice of purging technique may be dictated by hydrogeologic properties
(particularly depth to water table and hydraulic conductivity). a.
Equipment selection shall comply with construction and configuration
requirements specified in Table 4.1. b.
A clean protective covering may be placed around the wellhead during
purging activities. If this protective covering becomes soiled, ripped, etc. it
must be replaced prior to sampling. c.
The total amount of water must be recorded.
Therefore, the volume must be measured during the purged operation.
The amount may be determined by:
1.
Collecting the water in a graduated container (i.e. bucket); or
2.
Calculating volume based on pumping rate.
Note: the pumping rate may not be constant; take this into account. d.
Record the time that actual purging begins in the field records. e.
Purging is considered complete if any one of these criteria are
satisfied:
1.
three well volumes and subsequent stabilization of field parameters a.
Stabilization of field parameters is defined as "consecutive
readings within 5% taken at least five minutes apart".
b.
Even if field parameters have not stabilized after 5 well volumes,
purging is considered complete and sampling can begin.
2.
five well volumes (field parameters not monitored);
3.
at least one fully dry purge. a.
It has been suggested that one dry purge may not be adequate and a second
dry purge may be necessary. Theories
concerning purging the sandpack or aeration of sandpack will not be discussed
here. f.
Except for "low recovery" wells, all wells shall be sampled
within 6 hours of purging.
1.
"Low recovery" wells or wells that have been purged complete
dry may be sampled as soon as sufficient sample matrix is available or up to 10
hours after purging.
2.
Wells that have not recovered sufficiently within 10 hours of purging
should not be sampled. g.
Lanyards
1.
All lanyards must be securely fastened to downhole equipment (bailers,
pumps, etc.).
2.
Equipment construction and decontamination shall follow guidelines
discussed in Purging and Sampling Equipment above (4.2.5.3.d).
3.
Bailer lanyards must be handled such that they do not touch the ground
surface. h.
Low Hydraulic Conductivity Monitor Wells (i.e. wells that can be purged
dry)
1.
The most straightforward method for removing all of the stagnant water in
wells screened in low hydraulic conductivity formations is to install the pump
in the screen area and pump the well dry.
2.
Although this procedure may allow the atmosphere to contact the area of
the aquifer immediately surrounding the well screen, it is the best way to
ensure that all the stagnant water has been removed.
3.
If required, allow the well to recover and purge the well a second time. i.
High Hydraulic Conductivity Wells (i.e. wells that cannot be purged dry)
1.
For those wells with dedicated purging/sampling systems where the pump is
set in the screened area of the well, complete evacuation of the stagnant water
column may not be possible.
2.
The degree to which the stagnant water column can be replaced by fresh
aquifer water will be a function of the aquifer transmissivity and the number of
well volumes pumped (Barcelona et al., 1983).
3.
If in doubt, a short pump test or slug test may be performed on each
monitor well and the number of well volumes calculated to assure replacement of
the stagnant water. j.
In general, when nondedicated pumps that are used for purging, the
purging process should be done with the pump as near to the top of the water
column as possible to ensure that no stagnant water remains in the well above
the screen after purging. k.
Peristaltic Pump - One end of a length of new or pre-cleaned tubing shall
be attached to the pumphead flexible hose and the other end immersed no deeper
than one foot into the water column. l.
Centrifugal Pumps
1.
To minimize cross contamination while purging, fuel driven centrifugal
pumps must be placed at least 10 feet from the well head and downwind of the
well.
2.
Sampling gloves shall be worn and discarded after positioning the pump.
Hands should be washed and new gloves shall be put on prior to sampling.
3.
The length of suction hose should be situated such that the pump will be
withdrawing water from the top of the column.
4.
If the pump rate exceeds the recovery rate of the well then the hose
should be lowered into the well as needed to accommodate the drawdown.
5.
The suction hose must have a footvalve installed to prevent purge water
from re-entering the well. m.
Electric Submersible Pumps
1.
The pump should be set as near the top of the water column as possible to
ensure that all stagnant water in the casing is removed and to minimize the
contact area of the delivery hose with water column.
2.
If the pump rate exceeds the specific capacity of the well then the pump
must be lowered to accommodate the drawdown.
3.
If the pump has a controller, the flow rate may be adjusted to be equal
(or nearly) to the well's pumping capacity. n.
Bladder Pumps
1.
This equipment shall be operated strictly according to the
owners/operators manual or sample integrity and representativeness may be
suspect.
2.
After determining water level, position the controller/compressor away
from the well and downwind (if fuel powered compressor or generator).
3.
Attach tubing and lower the pump to a depth of 3 - 5 feet below the
surface of the water.
4.
If the pump is positioned too deep all of the stagnant water may not be
purged. If positioned too shallow
purging time will be slower as the bladder fills under standing head pressure.
5.
Adjust the pump position to follow the water level drawdown, if
necessary.
6.
It may be necessary to adjust the purging rate so that it is equivalent
to the drawdown rate.
7.
Discharge must be directed into graduated bucket or equivalent to
determine the number of well volumes. o.
BK Hand Pump
1.
For the B-K Pump, the intake is lowered to the top of the water column by
attaching additional 5-foot sections onto the pump.
2.
By changing the stroke of the actuating rod the pumping rate can be made
compatible with the well-specific yield. p.
Bailers
1.
The bailer must be handled carefully so as not to contaminate it prior to
use.
2.
The bailer must be lowered through the well and into the formation water
slowly. Allowing the bailer to drop
into the formation water with a splash is not acceptable.
3.
The bailer should be used to pull purge water from the top of the water
column so that fresh aquifer water can be pulled in through the screen.
This technique shall be performed until the requisite number of well
volumes have been evacuated. q.
All purging activities shall be documented in the field notes.
See Section 5.5 for the specific information that must be included. 4.2.5.6
Groundwater Sampling Techniques a.
Equipment Considerations
1.
Some pumps may be used for sampling groundwater.
All notes and restrictions as defined in Table 4.1 and discussed in
Purging and Sampling Equipment (Section 4.2.5.3) shall be followed when using
pumps to collect samples. a.
NOTE: The only pump that is currently approved for use in
collecting volatile samples is an all stainless steel and teflon bladder pump.
2.
Other than the actual sampling device, intermediate vessels should not be
used during the sample collection process.
This is especially true of any compound where loss of sample is a problem
(O&G, TRPH and VOCs). For all
trace compounds, the sample should come in contact with as few surfaces or
vessels as possible since excessive handling can result in contamination or
sample loss.
3.
Dedicated Sampling Equipment a.
The use of dedicated equipment is recommended since it significantly
reduces the chance of cross-contamination. b.
Dedicated is defined as equipment that is to be used solely for one
location for the life of that equipment (permanently mounted pump or permanently
dedicated bailer). Bringing 5
bailers on site to sample 5 monitor wells does not constitute dedication UNLESS
the bailers were purchased for the project, and each bailer is specifically
assigned to purge and/or sample a particular well. c.
All material construction and restrictions from Table 4.1 also apply to
dedicated equipment. Equipment
should be purchased with the most sensitive analyte of interest in mind. d.
Cleaning/Decontamination
1.
Dedicated pumps shall be cleaned prior to installation.
They need not be cleaned prior to each use but should be cleaned when
they are withdrawn for repair or servicing.
2.
Any permanently mounted tubing need not be cleaned.
3.
Any replaceable or temporary tubing shall be cleaned as specified in
Section 4.1.7.
4.
Equipment blanks on dedicated pumps shall be required when the tubing is
cleaned or replaced and shall be collected through that portion of the tubing
that is accessible.
5.
Dedicated bailers, if stored in the well, must be suspended above the
water column and completely decontaminated between sampling events. a.
After sampling is complete, they shall be rinsed with tap water and/or
analyte-free water, wrapped to prevent contamination, and stored on- or off-site
until the next sampling event. b.
The sampling equipment shall be decontaminated prior to on-site arrival
UNLESS the equipment is stored on-site. In
the latter case, the dedicated bailer shall be fully decontaminated prior to
on-site use. c.
A precleaned equipment blank shall be collected prior to reintroducing
the cleaned bailer into the water column. b.
Sampling with Bailer
1.
When a bailer is used for sampling, the integrity of the sample collected
is highly dependent upon the sampler's skill and familiarity with proper
sampling techniques.
2.
It is recommended that for a particular site only two persons perform
sampling to minimize personnel handling variation.
3.
Just prior to sampling, several bailer amounts of sample groundwater
shall be collected to rinse the bailer. a.
Discard the water appropriately (see Waste Disposal). b.
This should not be done if the analytes of interest include Oil &
Grease, TRPH, etc. (see Section 4.0.3). As
stated earlier, intermediate vessels or sampling equipment are never rinsed if
these compounds are to be sampled.
4.
All collection activities shall be done carefully so as to not stir up
any sediments.
5.
The following procedure describes general bailing techniques: a.
Field personnel should wear protective gloves (see Section 4.0.2). b.
Attach a fresh length of monofilament or braided nylon line to the
bailer. Alternately, a precleaned permanent lanyard may be used. c.
The bailer or lanyard must not be allowed to touch the ground during
purging or sampling. d.
Lower the bailer slowly and gently into contact with the water so that
agitation of the water column is minimized. e.
Attempt to sample from the same depth in the well each time, preferably
within or just above the screened zone of the well. f.
Do not allow the bailer to touch the bottom of the well so that bottom
sediment is incorporated into the sample. g.
Retrieve the bailer smoothly. Collecting
the lanyard between the thumbs of each hand seems to be the preferred method. h.
Discard the first few inches of water in the bailer and fill the
appropriate sample bottles so that a minimum of turbulence is created to avoid
aeration. I.
Discard the last few inches of water in the bailer. j.
Add preservatives (if necessary), check the pH of all pH-adjusted samples
(except VOCs). k.
Attach and/or complete the sample container labels, record information in
field notes, place samples on wet ice (if required) and protect all samples from
sun.
6.
Sampling with disposable bailers, though acceptable, is not recommended
as a standard procedure for environmental sampling. a.
Disposable bailers of the appropriate construction material are
available. High density polyethylene (HDPE) bailers are acceptable for
all inorganic parameters (and free product thickness). b.
Teflon bailers are also available as a disposable for use where organics
are of concern. c.
As the agency charged with solid waste management as well as
environmental sampling and analysis, the Department cannot encourage the use of
disposable equipment for all situations. In
situations where expensive, permanent sampling equipment may be destroyed or
damaged by sampling a concentrated waste, the used of disposable equipment is
recommended. d.
Precleaned equipment blanks are required for disposable equipment (see
Quality Control for frequency). c.
Sampling with Pumps
As a general rule, pumps shall not be used to
collect samples if organics are of interest.
There are two exceptions: 1)
use of the peristaltic pump with a trap (see Fig. 4.1 for specific
configuration) for EXTRACTABLE organics; and 2) use of an all Teflon and
stainless steel bladder pump for all organics.
1.
Peristaltic Pump a.
Organics
1.
Assemble the components of the pump according to the Fig. 4.1.
2.
The container shall be a glass or teflon bottle.
The sample container is recommended, however, if an intermediate vessel
is used, it shall be decontaminated between wells per Section 4.1.4.1.
3.
All equipment that contacts the groundwater BEFORE the sample container
shall be of Teflon, stainless steel or glass construction, including the
transport tubing to and from the sample container, the interior liner of the
container cap and all fittings. UNDER
NO CIRCUMSTANCES CAN A RUBBER STOPPER BE USED AS THE CAP.
4.
Connect the outflow tubing from the container to the influent side of the
peristaltic pump.
5.
Turn the pump on and allow the container to fill approximately 1/4 full.
6.
Turn the pump off, disconnect the container, rinse the bottle and discard
the contents.
7.
Repeat the process a second time. Note
restrictions on rinsing in Section 4.0.3.
8.
Turn pump on to fill the container.
9.
If an intermediate container is used, distribute the sample into
appropriate containers.
10.
If the sample container is used, discard a
small portion of the sample, to allow an air space.
11.
Preserve (if required), label and complete field notes. b.
Inorganics
1.
Inorganic samples may be collected from the effluent tubing, and there
are few restrictions on tubing type (see Table 4.1).
2.
If samples are collected from the pump, all tubing (including the tubing
in the head) shall be changed between wells.
2.
Bladder Pump a.
The flow rate shall be reduced after purging to a smooth, even flow. b.
When sampling for VOCs, the flow rate must be reduced to 500 ml/minute
(approx. 0.1 gallon/min).
3.
Other pump types a.
Sampling for INORGANICS ONLY may be conducted with most other pump types
(see Table 4.1). b.
The flow rate during sample collection shall be a smooth even flow. c.
All tubing and the pump shall be decontaminated between wells. d.
Sampling Wells that have Free-Phased Product
[[1. The
Department does not recommend the sampling of wells with free floating product
for trace contaminants. This
concerns primarily petroleum related sites, but includes any chemical product
(e.g. solvent) that floats on the water table.
Sampling is acceptable if the information is to be used for the purpose
of remedial design.
2.
Sample data from such wells cannot provide useful information regarding
the level of contamination. Furthermore,
the Department believes that these wells may never provide legitimate data as
they may have become (permanently) chemically damaged by the product being in
contact with the well casing for an extended period of time.
3.
The Department does reserve the right to require sampling of these wells,
not for levels of trace contaminants, but for confirmation of an appropriate
remediation technique. This type of sampling is performed BELOW the product layer
(see 4.2.5.6.f. below).]] e.
Free Product Sampling
1.
Free product is normally sampled for two reasons: a.
documentation for its existence (and thickness), as required in the Tanks
regulations; and b.
determination of the type of product so that the proper analyses can be
performed to determine extent. This
is only feasible for relatively recent releases as weathered product may not be
able to be identified.
2.
Free product may be evident in a cased monitor well or an open
excavation.
3.
It is recommended that plastic (acrylic, clear PVC) bailers be used for
sampling the monitor wells. Optionally,
disposable (HDPE) bailers are acceptable. Other
wide-mouth vessels may be used for sampling the excavation. a.
Monitoring well:
1.
If free product (defined in 17-770 as product in excess of 0.1 inch) or
product globules are present in a monitoring well, a precleaned bailer is used
to collect the sample.
2.
Once the bailer is withdrawn product thickness is measured to the closest
0.1 inch.
3.
A portion of the product is poured into a glass vial.
4.
As a concentrated waste, this sample must be wrapped to prevent breakage,
isolated from other samples, iced to 4 C, and proper chain of custody
information completed. b.
Excavation:
1.
If free product is observed in an open excavation, the glass sample
container or a precleaned intermediate vessel may be used to collect the sample.
2.
A lanyard (e.g. braided nylon) is tied to the container and lowered into
the excavation.
3.
Care must be taken not to introduce solid material into the container as
it is being lowered or retrieved.
4.
If sufficient water is available, a bailer can be used.
5.
Though not recommended, screened casing can be placed (or augured and
placed) in the bottom of the excavation and sampled with a bailer.
6.
Avoid dangerous situations, such as standing too close to the edge of an
excavation, riding in the backhoe bucket, or entering a trench or excavation
that may collapse.
7.
All applicable OSHA regulations should be followed.
4.
Equipment which is dedicated to sampling free product does not need to be
cleaned according to the standard, full decontamination protocols.
Acrylic or PVC bailers that are never used for trace contaminant sampling
may be cleaned as listed below. It
is recommended that all cleaning be done in the lab, office, or base of
operations and not in the field. a.
disassemble bailers and intermediate vessels and soak in hot, soapy tap
water using a brush to clean away all particulates and greasy films, b.
rinse with hot tap water, c.
thoroughly rinse with DI water d.
An optional acid rinse may be used to strip the equipment of any hard to
clean residues. e.
The solvent rinse is not mandatory since this equipment is not used for
contaminant sampling, other than the product itself.
It is not recommended on clear acrylic. f.
Sampling Below Product
1.
This type of depth-specific sampling is performed only at the request of
DER or its designee. An attempt is made to sample the dissolved constituents in
the water column below the product layer.
2.
These data provides information that helps define adequate groundwater
treatment. Without these data,
incorrect (and sometimes unnecessarily expensive) remediation techniques may be
designed for a situation where they are not required.
3.
There are some substantial logistical problems involved with sending a
sampler through free product to sample the groundwater below.
Although there are some products designed specifically for this type of
sampling, they are expensive and the results may not be commensurate with their
cost. The use of
"self-engineered" equipment or coverings may be the best option.
4.
These data are only to be used for qualitative use and will aid in
deciding on an appropriate remediation technique.
5. Wrapping
bailers and tubing in plastic seems to be the most popular technique in getting
past the product layer.
6.
Though not recommended, some have wrapped submersible pumps in several
layers of plastic and retrieved each layer by a separate lanyard.
7.
One suggestion would be to use a rigid piece of stainless steel tubing
wrapped in plastic. a.
Once the covered tubing is past the layer, pull up on the plastic,
piercing the plastic and exposing the (somewhat) clean tubing inlet. b.
Introduction of the wrapped hose must be done slowly so as to not entrain
any more product into the dissolved layer located below. c.
Also, this must be done with a peristaltic pump or a vacuum pump linked
to a trap bottle. To use this
setup, the water table must be no deeper than 15-20 feet, realizing that actual
sampling may be occurring several feet below the product layer. g.
Sampling Dissolved Metals
1.
In order to collect a "representative" sample for the purpose
of monitoring compliance with groundwater standards for metals, it may be
necessary to field filter a sample prior to preservation.
2.
In situations where the static level in the well allows use of a
peristaltic pump, the groundwater sample shall be pumped directly from the well
through an in-line filter. a.
A disposable, high capacity, .45 um filter is an acceptable filter for
most applications. See Fig. 4.2 and Table 4.1 for allowable equipment setups. b.
In field use, the filter must first be flushed with 30 - 50 mls of
deionized water or an inert gas to remove atmospheric oxygen. c.
The filter must be inserted on the high pressure side (i.e. on the
delivery side) of the peristaltic pump. VACUUM
FILTRATION IS NOT ACCEPTABLE. d.
The sample delivery tube must be long enough (greater than 2 feet) such
that back-diffusion of oxygen to the filter is negligible. e.
New or precleaned silastic tubing shall be installed in the pump at each
monitor well.
3.
In situations where the static water level in the well is too deep for a
peristaltic pump to be used directly, there are several alternatives: a.
Groundwater may be sampled with an appropriately constructed bailer.
The intake tube of the peristaltic pump is inserted into the full bailer
and water pumped through a filter as described above. b.
Any submersible pump of appropriate construction for which the flow rate
can be adjusted may be used for water levels below 20'-25'.
c.
Pressurized HDPE and Teflon bailers may also be used. d.
See the specific section concerning field filtration in Table 4.1 for all
acceptable alternatives.
4.
It is important that this operation is carried out as rapidly as possible
and in such a way that sample agitation and exposure to atmospheric oxygen is
minimized. It is for this reason
that pouring the sample into any intermediate vessel for subsequent filtration
IS NOT allowed. This includes barrel or syringe filters.
Once the sample is collected into a sample container, preservation and pH
checks should be completed. 4.2.6
Wells with In-Place Plumbing Wells with in-place plumbing are
generally encountered at wellfields, industrial facilities and private
residences. See separate
discussions below on sampling potable water wells. 4.2.6.1
Purging a.
The volume to be purged depends on several factors: the depth and
diameter of the well, whether the pumps are running continuously or
intermittently, how close to the source the sample can be collected, and the
presence of any storage/pressure tanks between the sampling point and the pump. b.
If storage/pressure tanks are present, an adequate volume must be purged
to totally exchange the volume of water in the tank (EPA, 1986). c.
Continuously Running Pumps
1.
If the pump runs continuously and the sample can be collected prior to a
storage/pressure tank, no purging is required, other than opening a valve and
allowing it to flush at maximum velocity for at least 15 minutes.
2.
If the pump runs continuously, and a storage/pressure tank is located
ahead of the sample location, the purge must include the entire storage volume
to ensure that a sample representative of the groundwater will be collected. d.
Intermittently Running Pumps
1.
If the pump runs intermittently it is necessary to determine the volume
to be purged, including storage pressure tanks that are located ahead of the
sampling location.
2.
The pump should then run continuously until the required volume has been
purged.
3.
When the well depth or diameter is unknown (as is frequently the case
with in-place plumbing) purging should be carried out by pumping the well for 15
minutes and until the pH, specific conductance and temperature stabilize. a.
In practice, stable sample chemistry is indicated when the purging
parameter values remain within 5% over two successive samples taken at least 5
minutes apart.
4.2.6.2 Sampling
All samples must be collected from the closest
spigot to the well head, with all screens or aerators removed, and with the flow
rate reduced to no more than 500 ml/min. The following procedures describe
generalized drinking water sampling from private potable wells.
If the samples are collected for compliance with the drinking water
regulations (Chapters 17-524, 17-550 or 17-555, F.A.C.), the samples must be
analyzed by a laboratory with Drinking Water Certification.
If the samples are being analyzed in response to other programs
(contamination assessment, consent order, etc.), the laboratory shall meet the
requirements of the specified Category.
4.2.7.1 General
Concerns a.
Appropriate containers and preservatives must be selected prior to
sampling.
1.
Containers and preservatives shall comply with Tables 4.2, 4.3, 4.4 and
4.5.
2.
Containers and preservatives may be obtained from a laboratory with
appropriate credentials (see discussion above).
3.
It is recommended that the laboratory add the appropriate preservative to
the container. b.
The laboratory may include special handling instructions with the sample
containers. These must be read carefully and must comply with the
generalized instructions listed below.
4.2.7.2 Sampling
Drinking Water Wells a.
As a general rule, purging and sample should be from a spigot closest to
the well head.
1.
If possible, the spigot should be before the holding tank and filters.
If this not possible, the contents of the holding tank must also be
purged.
2.
Remove all aerators and filters (if possible). b.
Depending on the running schedule of the well and the placement of the
pressure tank, purge the system as described in Section 4.2.6.1. c.
If the capacity of the pressure tank is not known, purge for at least 15
- 20 minutes at maximum velocity. d.
Reduce flow to approximately 500 ml/minute (a 1/8" stream). e.
Sample Containers with no preservatives:
1.
Remove the screw cap from the bottle.
Do not touch the interior of the cap or the container with hand or the
spigot.
2.
Fill approximately 1/4 full, rinse the interior of the container and
discard the water.
3.
DO NOT RINSE CONTAINERS IF collecting samples for oil and grease, total
recoverable hydrocarbons, volatile organics (including trihalomethanes) or
microbiologicals.
4.
Tilt the container so that flow falls onto the interior surface.
DO NOT AGITATE OR SHAKE CONTAINER WHILE FILLING.
5.
Fill the bottle to almost to capacity (if collecting VOC or
trihalomethane samples, see 4.2.7.2.i below).
6.
Replace the screw cap securely on the bottle. f.
Sample containers with preservatives.
1.
Follow the same protocol outlined above, deleting the rinse.
2.
Since some of the preservatives may react with the sample water, hold the
open end of the container away from you while filling.
3.
After replacing the cap, gently tip the container several times to mix
the preservatives. g.
Affix a sample label and seal (if required), and complete the
chain-of-custody form. h.
Place the sample bottle in a plastic sample bag and cool to 4 C on wet
ice. i.
Special Sampling Protocols
The special precautions for the types of
samples discussed in Section 4.2.2 shall be followed.
4.2.7.3 Sampling
Drinking Water Sources for Lead and Copper a.
Selection of the sampling point is dependent on whether the sample is
being taken to verify compliance with the Drinking Water Regulations.
If so, the sample must be collected from a COLD WATER tap in either the
kitchen or bathroom. b.
Samples must be collected after the water HAS NOT been used for at least
SIX HOURS. c.
DO NOT FLUSH OR PURGE THE SYSTEM. d.
Collect the first flush into the sample container for trace metals.
DO NOT RINSE SAMPLE CONTAINER. e.
Tilt the container so that the initial flow falls onto the interior
surface. DO NOT AGITATE. f.
If the container was prepreserved, hold the open end of the container
away from you while filling. g.
Add preservatives (if needed). h.
Replace screw cap and gently tip the container several times to mix the
preservatives. I.
Affix a sample label and seal (if required), and complete the
chain-of-custody form. j.
Place the sample bottle in a plastic sample bag. 4.2.8
Drinking Water Supply System Sampling The following protocols shall be
followed:
1. When
sampling for drinking water compliance, the sampling spigot is normally
designated by permit or municipal authorities.
The location may be near the supply line or may be an outside spigot on a
private residence.
2.
Procedures to sample drinking water directly from the supply system is
the same as above, except for treatment of residual chlorine. a.
Lines shall be flushed for 2 to 5 minutes before collecting any samples. b.
Reduce the flow rate to less than 500 ml/min (1/8" stream) before
collecting samples.
3. In
many instances, the water supply to residences maybe treated with chlorine which
may cause interference with certain types of analyses (ex: VOC; Semi-Volatiles
and some bacteriological samples). Residual
chlorine must be treated with the addition of sodium thiosulfate (Na2S2O3).
4. Utilizing
chemical kits (such as HACH), test the water in a separate container for
residual chlorine. If residual
chlorine is present, collect the sample in the appropriate sample container(s)
using the required preservatives. a.
Immediately upon sample collection add 0.008% Na2S2O3 or 100 mg of
Na2S2O3 per 1 liter of sample water directly into the sample container. b.
After replacing the cap, tip the container several times to mix the
preservative.
5. Affix
a sample label, seal and transport on wet ice.
6. Lead
and copper shall be sampled according to protocols outlined in 4.2.7.3. Temporary well points include those
drilled with augers as well as those pushed with "direct push" or DPT
devices. These types of wells are
not permanently installed. 4.2.9.1
Use a.
Temporary well points may be used for PRELIMINARY INVESTIGATIONS and as a
SCREENING TOOL. [[b.
For formal site work (not preliminary or PCAP), temporary well points may
only be allowed under emergency situations.
These are:
1.
DOT right-of-ways,
2.
private property where a permanent well cannot be placed, or
3.
inside or up against a structure.]] c.
DER will determine whether temporary well points are warranted. d.
If these wells are used to provide formal data, these restrictions apply:
1.
Use precleaned equipment as described in Table 4.1;
2.
Well must be purged of 3-5 well volumes (or dry);
3.
Sampling with a peristaltic pump a.
Extractable organics shall be collected via an all-Teflon and -glass
organic trap configuration (see Figure 2.1); b.
VOCs shall not be collected through a pump, but the Teflon pump tubing is
allowed to fill via ambient pressure, capped with stopper or gloved finger,
carefully withdrawn from the well, and drained into appropriate vials. c.
Refer to protocols listed in 4.2.5.5 and 4.2.5.6 for specific information
on sampling and configuration.
4.
Sampling with bailers a.
In some cases, sampling may be accomplished with a 3/4" bailer. b.
All equipment construction restrictions shall be followed. c.
Refer to bailer sampling protocols in section 4.2.6.5. 4.2.10
Airstripper and Remedial Treatment System Sampling a.
Collect effluent samples from airstripper units in a similar manner to
those described for Drinking Water Supply Systems (Section 4.2.8). b.
Remove any tubing from the sampling port and flush for one to two
minutes. c.
Reduce flow rate to less than 500 ml/min. and begin sample collection. 4.2.11
Bioassay Sampling When collecting samples for bioassays,
the sampling protocols outlined in Section 4.2.3 (Surface Water) and 4.2.4
(Wastewater) shall be followed. The holding time for bioassay samples is 72 hours. |