|
Site index
Home
Created by:

| |
Preventive maintenance is the key ingredient to possessing analytical
equipment that will produce reliable data over the life of the instrument.
Responsibility for preventive maintenance lies with the analyst and
supervisory personnel in charge of the monitoring equipment. The analytical
staff must be dedicated to the implementation of the preventive
maintenance program and always watchful for signs that there is a need
for maintenance activities. The analyst and supervisory personnel must be
supported by vendor specialists or in-house experts that handle activities
beyond simple repairs or maintenance.
The Preventive Maintenance Program must consist of:
1. A written PM
schedule;
2.
Documentation of all maintenance and repairs (records must be kept in an
easily accessible manner);
3. Vendor
operation and maintenance manuals available for all instrumentation; and
4. A
written contingency plan specifying that backup equipment will be maintained for
all instrumentation or stating that sampling events will be postponed and
current sample load be invalidated until repairs are accomplished.
If samples are sent to another laboratory the subject laboratory must
have an approved CompQAP for the parameters of concern and the DER Project
Manager must be notified if the analytical work is being performed under a
Quality Assurance Project Plan.
Table 8.1 identifies preventive maintenance activities by instrument type
required by DER with recommended frequencies.
Please note that it may be necessary to perform activities more
frequently depending on heavy workloads, sample types analyzed and/or instrument
performance. If the instrument manufacturer recommends more frequent or
additional maintenance activities these shall also be incorporated into the
facility maintenance program.
Table
8.1
PREVENTIVE
MAINTENANCE ACTIVITIES
|
INSTRUMENT/ACTIVITY
|
|
|
FREQUENCY
|
|
|
|
|
|
|
|
|
AA SPECTROPHOTOMETER
(FLAME)
|
|
|
|
|
|
|
|
Clean
nebulizer
|
SA
|
|
|
|
|
Clean
spectrophotometer quartz windows
|
W
|
|
|
|
|
Burner
head cleaned; check tubing, pump and lamps
|
D(1)
|
|
|
|
|
O
rings checked
|
M(1)
|
|
|
|
|
Fine
Tune Wavelength; Check optics
|
A
|
|
|
|
|
Check
electronics
|
A(3,4)
|
|
|
|
|
|
|
|
|
AA SPECTROPHOTOMETER
(FURNACE)
|
|
|
|
|
|
|
Check
graphite tubes
|
D(1)
|
|
|
|
|
Flush
autosampler tubing
|
D
|
|
|
|
|
Replace
graphite electrodes and Shrouds
|
SA
|
|
|
|
|
Clean
furnace housing and injector tip
|
D
|
|
|
|
|
Check
electronics
|
A(3,4)
|
|
|
|
|
|
|
|
|
AA SPECTROPHOTOMETER
(COLD VAPOR & HYDRIDE)
|
|
|
|
|
|
|
|
Flush
tubing (automated systems)
|
D(1)
|
|
|
|
|
Check
absorption cell for vitrification
|
D(1)
|
|
|
|
|
Replace
or clean quartz cell
|
3
|
|
|
|
|
Check
electronics
|
A(3,4)
|
|
|
|
|
|
|
|
|
ICP
|
|
|
|
|
|
|
|
Clean
and realign torch
|
M
|
|
|
|
|
Clean
nebulizer and spray chamber; Check peristaltic
|
W(1)
|
|
|
|
|
pump
tubing and vacuum pump oil
|
|
|
|
|
|
Check
entire optical system (mirrors, windows, etc.)
|
A(3,4)
|
|
|
|
|
Check
water lines, torch compartment and gases
|
D
|
|
|
|
|
Check
electronics (voltages, waveforms, etc.)
|
SA
|
|
|
|
|
Check
wavelength calibration and adjust as needed
|
SA
|
|
|
|
|
Run
interference (interelement) standard
|
SA
|
|
|
|
|
|
|
|
|
GAS CHROMATOGRAPHS
|
|
|
|
|
|
|
GENERAL
|
|
|
|
|
|
|
Check
septa, cylinder gas pressure,
|
D
|
|
|
|
|
oxygen/moisture
traps
|
|
|
|
|
|
Bake
out injector body
|
2
|
|
|
|
|
Check
electronics (voltages, waveforms, etc.)
|
Q(3,4)
|
|
|
|
|
Check
GC temperature calibrations (injector, oven, detector)
|
Q
|
|
|
|
COLUMNS
|
|
|
|
|
|
|
Change
glass sleeve inserts, shorten ends of columns, change glass wool plugs,
check for leaks or replace
|
3
|
|
|
|
ELECTRON CAPTURE
DETECTOR
|
|
|
|
|
|
|
Wipe
Tests
|
SA
|
|
|
|
|
Hydrogen
cleaning
|
3
|
|
|
|
|
Returned
to factory for cleaning and refoil
|
3,4
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Table
8.1, cont.
PREVENTIVE
MAINTENANCE ACTIVITIES
|
INSTRUMENT/ACTIVITY
|
|
|
FREQUENCY
|
|
GAS CHROMATOGRAPHS,
Cont.
|
|
|
|
|
|
|
FLAME IONIZATION
DETECTOR
|
|
|
|
|
|
|
Clean
|
Q
|
|
|
|
Replace
Flame Tip
|
A
|
|
|
PHOTO IONIZATION
DETECTOR
|
|
|
|
|
|
|
Clean
lamp
|
M(1)
|
|
|
HALL ELECTROLYTIC
CONDUCTIVITY DETECTOR
|
|
|
|
|
|
|
Replace
resin, change solvent and clean conductivity cell
|
3,4
|
|
|
|
Change
Ni tube
|
Q
|
|
|
NITROGEN PHOSPHORUS
DETECTOR
|
|
|
|
|
|
|
Clean
|
Q
|
|
|
MASS SPECTROMETER
|
|
|
|
|
|
|
Replace
vacuum pump oil and change desiccant
|
A
|
|
|
|
Check
ion source and analyzer (dismantle and clean, replace parts as needed)
|
Q
|
|
|
|
Check
mechanicals (vacuum pumps, relays, gas pressures and flows)
|
Q
|
|
|
|
Check
mass calibration w/ FC-43 (perfluorotributyl- amine)
|
D
|
|
|
PURGE AND TRAP
|
|
|
|
|
|
|
Clean
sparger
|
W
|
|
|
|
Change
Trap
|
A
|
|
|
|
Bake
Trap
|
2
|
|
|
|
Check
purge flow
|
M
|
|
|
|
Check
for leaks
|
M
|
|
|
|
|
|
|
|
HIGH PRESSURE LIQUID
CHROMATOGRAPHY
|
|
|
|
|
|
|
|
Gas
lines checked for leaks
|
D
|
|
|
|
Clean
mobile phase flow system with nitric acid
|
SA
|
|
|
|
Clean
detector flow cells with nitric acid
|
SA
(3)
|
|
|
|
Clean
injection valve
|
A
|
|
|
|
Check
solvent filters
|
W
|
|
|
|
Check
pumps seals and check valve assemblies (clean and replace as pressures
& flows of mobile phase indicate)
|
D
|
|
|
|
Lubricate
oil felts, if present
|
M
|
|
|
|
Lubricate
post column reagent pumps and check valve assembly oil felts
|
M
|
|
|
|
|
|
|
INFRARED
SPECTROPHOTOMETER
|
|
|
|
|
|
|
|
Clean
instrument housing
|
M
|
|
|
|
Change
desiccant and clean cells
|
Q
|
|
|
|
Clean
windows
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Table
8.1, cont.
PREVENTIVE
MAINTENANCE ACTIVITIES
|
INSTRUMENT/ACTIVITY
|
|
|
FREQUENCY
|
|
|
AUTOANALYZERS
|
|
|
|
|
|
|
|
Check
for leaks, flush out system and clean up spills after use
|
D
|
|
|
|
|
Clean
sample probe and check all tubing for wear and discoloration
|
M(1)
|
|
|
|
|
Clean
optics
|
Q
|
|
|
|
|
Oil
sample motor, lubricate gears, clean flow cell
|
SA
|
|
|
|
|
Clean
pump rollers, platens and colorimeter filters
|
M
|
|
|
|
|
|
|
|
|
ION CHROMATOGRAPH
|
|
|
|
|
|
|
|
Check
for leaks
|
D
|
|
|
|
|
Check
all lines for wear and discoloration
|
W(1)
|
|
|
|
|
Check
pump pistons
|
A
|
|
|
|
|
|
|
|
|
TOC ANALYZER
|
|
|
|
|
|
|
|
Change
injection needle, clean injection port,
|
M
|
|
|
|
|
and
change catalyst
|
|
|
|
|
|
Inspect
combustion tube
|
SA
|
|
|
|
|
|
|
|
|
RADIOCHEMISTRY
|
|
|
|
|
|
|
LOW BACKGROUND GAS
PROPORTIONAL COUNTERS
|
|
|
|
|
|
|
Clean
sample drawers and windows
|
W
|
|
|
|
LUCAS CELLS
|
|
|
|
|
|
|
Clean
Windows
|
M
|
|
|
|
|
|
|
|
|
REFRIGERATORS,
INCUBATORS, OVENS
|
|
|
|
|
|
|
|
Clean
interior
|
M
|
|
|
|
|
Check
thermometer temperature against certified thermometer or equivalent
|
A
|
|
|
|
|
|
|
|
|
ANALYTICAL BALANCES
|
|
|
|
|
|
|
Clean
pan and compartment
|
D
|
|
|
|
|
Check
with class S weights
|
M
|
|
|
|
|
Manufacturer
cleaning and calibration
|
A
|
|
|
|
|
|
|
|
|
AUTOCLAVES
|
|
|
|
|
|
|
|
Gaskets
checked
|
W(1)
|
|
|
|
|
Timing
mechanism checked
|
SA
|
|
|
|
|
Clean
interior
|
M
|
|
|
|
|
Sterilization
indicator tape
|
D
|
|
|
|
|
|
|
|
|
MICROSCOPES
|
|
|
|
|
|
|
|
Clean
optics
|
M
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Table
8.1, cont.
PREVENTIVE
MAINTENANCE ACTIVITIES
|
INSTRUMENT/ACTIVITY
|
|
|
FREQUENCY
|
|
UV/VIS SPECTROPHOTOMETER
|
|
|
|
|
|
|
|
Lamp
alignment checked
|
3,4
|
|
|
|
Windows
cleaned
|
M
|
|
|
|
Check
and adjust photomultiplier sensitivity and wavelength resolution
|
A
|
|
|
|
Replace
lamp
|
4
|
|
|
|
Clean
sample compartment before and after each use
|
D
|
|
|
|
Check
electronics
|
A(3,4)
|
|
|
|
Adjust
baseline for smoothness through entire wavelength range
|
SA
|
|
|
|
Clean
cuvettes after each use
|
D
|
|
|
|
|
|
|
TOX ANALYZER (TOTAL
ORGANIC HALOGENS)
|
|
|
|
|
|
|
|
Clean
titration cell; clean inlet and exit tube
|
W
|
|
|
|
Clean
pyrolysis tube, recoat electrodes
|
5
|
|
|
|
|
|
|
|
pH AND ION SELECTIVE
ELECTRODES
|
|
|
|
|
|
|
PROBE
|
|
|
|
|
|
Check
probe for cracks and proper levels of filling solution; check reference
junction; clean electrode
|
D(1)
|
|
|
|
Check
response time
|
D
|
|
|
METER
|
|
|
|
|
|
Check
batteries and electronics for loose connections and cracked leads
|
D(1)
|
|
|
|
|
|
|
TURBIDIMETER
|
|
|
|
|
|
|
Clean
instrument housing
|
M
|
|
|
|
Clean
cells
|
D
|
|
|
|
|
|
|
CONDUCTIVITY METER
|
|
|
|
|
|
|
Check
batteries and probe cables
|
D
|
|
|
|
Replatinize
Probe
|
5
|
|
|
|
|
|
|
DISSOLVED OXYGEN
METERS
|
|
|
|
|
|
PROBE
|
|
|
|
|
|
Check
membrane for deterioration; check filling solution
|
D(1)
|
|
|
METER
|
|
|
|
|
|
Battery
level and electronics checked
|
D(1)
|
|
|
|
|
|
|
THERMOMETERS
|
|
|
|
|
|
|
Check
for cracks and gaps in the mercury
|
D(1)
|
|
|
|
|
|
|
TEMPERATURE PROBE
|
|
|
|
|
|
|
Check
connections, cables
|
D
|
|
|
|
Check
against calibrated thermometer
|
D
|
|
|
|
|
|
|
AUTOSAMPLERS
|
|
|
|
|
|
|
Check
needles and tubing
|
D(1)
|
|
|
|
Clean
|
Q
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Table
8.1, cont.
PREVENTIVE
MAINTENANCE ACTIVITIES
|
INSTRUMENT/ACTIVITY
|
|
|
FREQUENCY
|
|
|
DATA SYSTEMS
|
|
|
|
|
|
|
|
Clean
computers, check battery backup and check
|
Q
|
|
|
|
|
ventilation
fans
|
|
|
|
|
|
|
|
|
|
AUTOMATIC SAMPLE
COLLECTION SYSTEMS
(ex. ISCO, Sigma, etc.)
|
|
|
|
|
|
|
Check
sampler operation (forward, reverse, automatic
|
D(6)
|
|
|
|
|
through
three cycles of the purge-pump-purge cycle)
|
|
|
|
|
|
Check
purge-pump-purge cycle when sampler is installed
|
D(7)
|
|
|
|
|
Check
the flow pacer that activates the sampler to assure proper operation
|
|
|
|
|
|
Check
desiccant
|
D(1,6)
|
|
|
|
|
Check
batteries
|
D(1,6)
|
|
|
|
|
Check
pumping rate against manufacturer's specifications
|
D(1,6)
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
KEY:
|
1
|
Replace
as necessary
|
|
D
|
daily*
|
|
2
|
High
background
|
|
W
|
weekly
|
|
3
|
Loss
of sensitivity or failing resolution
|
|
M
|
monthly
|
|
4
|
Erratic
response
|
|
Q
|
quarterly
|
|
5
|
QC
failure
|
|
SA
|
semi-annually
|
|
6
|
Prior
to sampling event
|
|
A
|
annually
|
|
7
|
In
situ (under field conditions)
|
|
|
|
*Daily
is defined as prior to use or a 12-hour period if equipment is run continuously
|